DICOM PS3.16 2021e - Content Mapping Resource

OB-GYN Report Templates 

TID 5000 OB-GYN Ultrasound Procedure Report

This is the Template for the root of the content tree for the OB-GYN ultrasound procedure report.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 5000. OB-GYN Ultrasound Procedure Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

BCID 12024 “OB-GYN Ultrasound Report Document Titles”

1

M

Root node

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

U

2b

>

HAS ACQ CONTEXT

CODE

EV (125203, DCM, "Acquisition Protocol")

1-n

U

DCID 12025 “OB-GYN Ultrasound Beam Path”

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

M

4

>

CONTAINS

INCLUDE

DTID 5001 “OB-GYN Patient Characteristics”

1

U

5

>

CONTAINS

CONTAINER

DT (111028, DCM, "Image Library")

1

U

6

>>

CONTAINS

IMAGE

1-n

M

7

>

CONTAINS

INCLUDE

DTID 5002 “OB-GYN Procedure Summary Section”

1

U

8

>

CONTAINS

INCLUDE

DTID 5004 “Fetal Biometry Ratio Section”

1-n

U

9

>

CONTAINS

INCLUDE

DTID 5005 “Fetal Biometry Section”

1-n

U

10

>

CONTAINS

INCLUDE

DTID 5006 “Fetal Long Bones Section”

1-n

U

11

>

CONTAINS

INCLUDE

DTID 5007 “Fetal Cranium Section”

1-n

U

12

>

CONTAINS

INCLUDE

DTID 5009 “Fetal Biophysical Profile Section”

1-n

U

13

>

CONTAINS

INCLUDE

DTID 5011 “Early Gestation Section”

1-n

U

14

>

CONTAINS

INCLUDE

DTID 5010 “Amniotic Sac Section”

1

U

15

>

CONTAINS

INCLUDE

DTID 5015 “Pelvis and Uterus Section”

1

U

16

>

CONTAINS

INCLUDE

DTID 5012 “Ovaries Section”

1

U

17

>

CONTAINS

INCLUDE

DTID 5013 “Follicles Section”

1

U

$Laterality = EV (7771000, SCT, "Left")

$Number = EV (11879-4, LN, "Number of follicles in left ovary")

18

>

CONTAINS

INCLUDE

DTID 5013 “Follicles Section”

1

U

$Laterality = EV (24028007, SCT, "Right")

$Number = EV (11880-2, LN, "Number of follicles in right ovary")

19

>

CONTAINS

CONTAINER

EV (59776-5, LN, "Findings")

1-n

U

20

>>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

EV (51852003, SCT, "Embryonic Vascular Structure")

21

>>

CONTAINS

INCLUDE

DTID 5025 “OB-GYN Fetal Vascular Ultrasound Measurement Group”

1

M

$AnatomyGroup = DCID 12141 “Fetal Vasculature Anatomical Location”

22

>

CONTAINS

CONTAINER

EV (59776-5, LN, "Findings")

1-n

U

23

>>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

EV (281496003, SCT, "Pelvic Vascular Structure")

24

>>

CONTAINS

INCLUDE

DTID 5026 “OB-GYN Pelvic Vascular Ultrasound Measurement Group”

1

M

$AnatomyGroup = DCID 12140 “Pelvic Vasculature Anatomical Location”


Content Item Descriptions

Row 6

No purpose of reference is specified.

DICOM PS3.16 2021e - Content Mapping Resource