DICOM PS3.16 2026a - Content Mapping Resource

Waveform Annotation Templates 

TID 3750 Waveform Annotations

This Root Template encodes a list of annotations for waveform data consisting of measurements or observations added at recording time or later provided either by a human reviewer (such as a cardiologist, neurologist, or technologist) or by an automated analysis algorithm.

Type:

Extensible

Order:

Non-Significant

Root:

Yes

Table TID 3750. Waveform Annotations

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

BCID 3048 “Waveform Annotations Document Title ”

1

M

Root node

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

U

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

M

4

>

HAS CONCEPT MOD

CODE

EV (130871, DCM, "Procedure annotated")

1-n

U

BCID 3670 “ECG Procedure Type”

BCID 3049 “EEG Procedure ”

5

>

HAS OBS CONTEXT

CODE

EV (1185780006, SCT, "Relative Time")

1

U

DCID 12102 “Temporal Period Relating to Procedure or Therapy”

6

>

CONTAINS

INCLUDE

DTID 3754 “Waveform Library”

1

U

7

>

CONTAINS

CONTAINER

EV (130870, DCM, "Waveform Annotations")

1

M

8

>>

HAS CONCEPT MOD

INCLUDE

DTID 4019 “Algorithm Identification”

1

U

9

>>

CONTAINS

CONTAINER

EV (130872, DCM, "Waveform Annotation Group")

1-n

M

10

>>>

HAS OBS CONTEXT

NUM

EV (130873, DCM, "Waveform Annotation Group Number")

1

M

UNITS = (1, UCUM, "no units")

11

>>>

HAS OBS CONTEXT

TEXT

EV (130874, DCM, "Waveform Annotation Group Label")

1

U

12

>>>

CONTAINS

INCLUDE

DTID 3751 “Waveform Pattern or Event”

1-n

U

$AnnotationClassification = EV (130860, DCM, "Pattern Event")

$AnnotationCode = BCID 3038 “Pattern Event”

13

>>>

CONTAINS

INCLUDE

DTID 3751 “Waveform Pattern or Event”

1-n

U

$AnnotationClassification = EV (130861, DCM, "EEG Annotation")

$AnnotationCode = BCID 3035 “EEG Annotation – Neurophysiologic Enumeration”

14

>>>

CONTAINS

INCLUDE

DTID 3751 “Waveform Pattern or Event”

1-n

U

$AnnotationClassification = EV (130862, DCM, "EMG Annotation")

$AnnotationCode = BCID 3036 “EMG Annotation – Neurophysiological Enumeration”

15

>>>

CONTAINS

INCLUDE

DTID 3751 “Waveform Pattern or Event”

1-n

U

$AnnotationClassification = EV (130863, DCM, "EOG Annotation")

$AnnotationCode = BCID 3037 “EOG Annotation – Neurophysiological Enumeration”

16

>>>

CONTAINS

INCLUDE

DTID 3751 “Waveform Pattern or Event”

1-n

U

$AnnotationClassification = EV (130864, DCM, "Device-related and Environment-related Event")

$AnnotationCode = BCID 3039 “Device-related and Environment-related Event”

17

>>>

CONTAINS

INCLUDE

DTID 3751 “Waveform Pattern or Event”

1-n

U

$AnnotationClassification = EV (130865, DCM, "Patient Consciousness")

$AnnotationCode = BCID 3050 “Patient Consciousness ”

18

>>>

CONTAINS

INCLUDE

DTID 3751 “Waveform Pattern or Event”

1-n

U

$AnnotationClassification = EV (130866, DCM, "ECG Annotation")

$AnnotationCode = BCID 3335 “ECG Annotation”

19

>>>

CONTAINS

INCLUDE

DTID 3752 “Waveform Measurement”

1-n

U

$Measurement = BCID 3040 “EEG Annotation - Neurological Monitoring Measurement”

20

>>>

CONTAINS

INCLUDE

DTID 3753 “Annotation Note”

1-n

U


Content Item Descriptions

Row 4

A coded descriptor of the sort of procedure the annotations apply to.

Row 5

Indicates the point in time when the annotations have been made relative to the waveform recording procedure.

Row 6

The Waveform Library provides potentially relevant characteristics of the waveform objects associated with the annotations. There is no requirement to include all, or any, of the waveform objects referenced in the annotations and measurements elsewhere in this template. The template may also include waveform objects that are associated with, but not directly referenced in, the annotations and measurements.

The Waveform Library is not replicating the content of the SOP Instance Reference Macro.

Row 10

Defines an identifier for a group of annotations analogously to Annotation Group Number (0040,A180) in Section C.10.10.1.4 in PS3.3 , which may be used, for example, for display purposes.

The number itself is not semantically significant, no ordering is required.

Row 11

A descriptive label for a group of annotations, e.g., to be used for display purpose.

DICOM PS3.16 2026a - Content Mapping Resource