Vascular Ultrasound Report Templates 

TID 5100 Vascular Ultrasound Report

This is the Template for the root of the Content Tree for the vascular ultrasound procedure report.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 5100. Vascular Ultrasound Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

BCID 12100 “Vascular Ultrasound Report Document Title”

1

M

2

>

HAS OBS CONTEXT

CODE

EV (307152002, SCT, "Temporal periods Relating to Procedure")

1

U

DCID 12102 “Temporal Period Relating to Procedure or Therapy”

3

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

U

4

>

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

M

4b

>

CONTAINS

INCLUDE

DTID 1025 “Indications for Procedure”

1

U

5

>

CONTAINS

INCLUDE

DTID 5101 “Vascular Patient Characteristics”

1

U

6

>

CONTAINS

CONTAINER

EV (111028, DCM, "Image Library")

1

U

7

>>

CONTAINS

IMAGE

1-n

M

8

>

CONTAINS

INCLUDE

DTID 5102 “Vascular Procedure Summary Section”

1

U

9

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (281231009, SCT, "Blood Vessel of Head")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12105 “Intracranial Cerebral Vessel”

10

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (281231009, SCT, "Blood Vessel of Head")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12105 “Intracranial Cerebral Vessel”

11

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (281231009, SCT, "Blood Vessel of Head")

$SectionLaterality = EV (66459002, SCT, "Unilateral")

$Anatomy = DCID 12106 “Intracranial Cerebral Vessel (Unilateral)”

12

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (119568004, SCT, "Artery of neck")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12104 “Extracranial Artery”

$AnatomyRatio = DCID 12123 “Carotid Ratio”

13

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (119568004, SCT, "Artery of neck")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12104 “Extracranial Artery”

$AnatomyRatio = DCID 12123 “Carotid Ratio”

14

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (70791007, SCT, "Artery of Lower Extremity")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12109 “Lower Extremity Artery”

15

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (70791007, SCT, "Artery of Lower Extremity")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12109 “Lower Extremity Artery”

16

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (122774002, SCT, "Vein of Lower Extremity")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12110 “Lower Extremity Vein”

17

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (122774002, SCT, "Vein of Lower Extremity")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12110 “Lower Extremity Vein”

18

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (75531005, SCT, "Artery Of Upper Extremity")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12107 “Upper Extremity Artery”

19

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (75531005, SCT, "Artery Of Upper Extremity")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12107 “Upper Extremity Artery”

20

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (122775001, SCT, "Vein Of Upper Extremity")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12108 “Upper Extremity Vein”

21

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (122775001, SCT, "Vein Of Upper Extremity")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12108 “Upper Extremity Vein”

22

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (303402001, SCT, "Vascular Structure Of Kidney")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12115 “Renal Vessel”

$AnatomyRatio = DCID 12124 “Renal Ratio”

23

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (303402001, SCT, "Vascular Structure Of Kidney")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12115 “Renal Vessel”

$AnatomyRatio = DCID 12124 “Renal Ratio”

24

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (118634008, SCT, "Artery of Abdomen")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12111 “Abdominopelvic Artery (Paired)”

25

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (118634008, SCT, "Artery of Abdomen")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12111 “Abdominopelvic Artery (Paired)”

26

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (118634008, SCT, "Artery of Abdomen")

$SectionLaterality = EV (66459002, SCT, "Unilateral")

$Anatomy = DCID 12112 “Abdominopelvic Artery (Unpaired)”

27

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (84421000, SCT, "Vein of Abdomen")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12113 “Abdominopelvic Vein (Paired)”

28

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (84421000, SCT, "Vein of Abdomen")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12113 “Abdominopelvic Vein (Paired)”

29

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (84421000, SCT, "Vein of Abdomen")

$SectionLaterality = EV (66459002, SCT, "Unilateral")

$Anatomy = DCID 12114 “Abdominopelvic Vein (Unpaired)”

29b

>

CONTAINS

INCLUDE

DTID 5103 “Vascular Ultrasound Section”

1

U

$SectionScope = DT (846654004, SCT, "Abdominal and pelvic vessels")

$Anatomy = DCID 12125 “Abdominopelvic Vessel”

30

>

CONTAINS

INCLUDE

DTID 5105 “Ultrasound Graft Section”

1

U


Content Item Descriptions

Row 1

Previously, an Enumerated Value of (125100, DCM, "Vascular Ultrasound Procedure Report") was specified for the Concept Name. This code has been retired in favor of codes from external schemes, as well as permitting alternative codes to be used for more specific reports, e.g., for a Carotid Artery report. See PS3.16 2020a.

Row 7

No purpose of reference is specified.