DICOM PS3.16 2026c - Content Mapping Resource

TID 5026 OB-GYN Pelvic Vascular Ultrasound Measurement Group

This Template is an anatomy specific container of OB-GYN pelvic vascular measurements.

Table TID 5026. Parameters

Parameter Name

Parameter Usage

$AnatomyGroup

The concept name of the vascular anatomy


Type:

Extensible

Order:

Significant

Root:

No

Table TID 5026. OB-GYN Pelvic Vascular Ultrasound Measurement Group

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

$AnatomyGroup

1

M

2

>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

MC

IFF anatomy has laterality

DCID 244 “Laterality”

3

>

HAS CONCEPT MOD

TEXT

EV (112050, DCM, "Anatomic Identifier")

1

U

4

>

CONTAINS

NUM

DCID 12119 “Vascular Ultrasound Property”

1-n

M

4b

>>

INCLUDE

DTID 301 “Measurement Content”

1

M

$Derivation = DCID 3627 “Measurement Type”

5

>>

HAS CONCEPT MOD

CODE

EV (125105, DCM, "Measurement Orientation")

1

U

DCID 12118 “Measurement Orientation”

6

>>

HAS PROPERTIES

NUM

EV (125106, DCM, "Doppler Angle")

1

U

UNIT = EV (deg, UCUM, "deg")

7

>>

HAS PROPERTIES

NUM

EV (125107, DCM, "Sample Volume Depth")

1

U

UNIT = EV (cm, UCUM, "cm")


Content Item Descriptions

Row 1

Specifies the anatomical context of the observations in the group.

Row 3

Differentiates between multiple structures such as the two umbilical arteries.

DICOM PS3.16 2026c - Content Mapping Resource