DICOM PS3.16 2026d - Content Mapping Resource

TID 5010 Amniotic Sac Section

This Template specifies a container for amniotic sac quadrant diameters and a derived index.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 5010. Amniotic Sac Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

DT (70847004, SCT, "Amniotic Sac")

2a

>

HAS OBS CONTEXT

INCLUDE

DTID 1008 “Subject Context, Fetus”

1

MC

IF this Template is invoked more than once

3

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = DT (11627-7, LN, "Amniotic Fluid Index")

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

4

U

$Measurement = DCID 12008 “OB-GYN Amniotic Sac Measurement”


Content Item Descriptions

Row 2a

This row shall be used to identify the fetus being described if there is more than one fetus being measured.

Row 3

The sum of the 4 quadrant diameters.

Row 4

The four amniotic sac quadrant diameters.

DICOM PS3.16 2026d - Content Mapping Resource