DICOM PS3.16 2026d - Content Mapping Resource

TID 5007 Fetal Cranium Section

The Fetal Cranium Template is a container for groups of biometric data of the fetal cranium.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 5007. Fetal Cranium Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (125004, DCM, "Fetal Cranium")

1

M

2

>

HAS OBS CONTEXT

INCLUDE

DTID 1008 “Subject Context, Fetus”

1

MC

IF this Template is invoked more than once to describe more than one fetus

3

>

CONTAINS

INCLUDE

DTID 5008 “Fetal Biometry Group”

1-n

M

$BiometryType = MemberOf {DCID 12007 “Fetal Cranium Measurement”

$TargetSite = DCID 12022 “Fetal Cranium Anatomic Site”


Content Item Descriptions

Row 3

The group of measurements. Only one group per biometry type.

DICOM PS3.16 2026d - Content Mapping Resource