DICOM PS3.16 2026a - Content Mapping Resource

TID 4303 Overall Prostate Finding

Summary of the overall assessment of the prostate gland, peripheral and transitional zone in separate container instantiations. Prostate gland assessment must include size measurements.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 4303. Overall Prostate Finding

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130554, DCM, "Overall Prostate Finding")

1

M

2

>

HAS OBS CONTEXT

TEXT

DT (112039, DCM, "Tracking Identifier")

1

M

3

>

HAS OBS CONTEXT

UIDREF

EV (112040, DCM, "Tracking Unique Identifier")

1

M

4

>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1

M

EV (255503000, SCT, "Entire")

5

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding site")

1

M

EV (41216001, SCT, "Prostate")

6

>

CONTAINS

INCLUDE

DTID 3909 “Best Illustration of Findings”

1-n

U

7

>

CONTAINS

INCLUDE

DTID 1410 “Planar ROI Measurements and Qualitative Evaluations”

1-n

MC

IF Rows 8, 9 are absent

$Measurement = BCID 218 “Quantitative Image Feature”

$Units = BCID 7181 “Abstract Multi-dimensional Image Model Component Unit”

$Derivation = BCID 7464 “General Region of Interest Measurement Modifier”

$Method = BCID 6147 “Response Criteria”

$FindingType = Value of Row 4

$TargetSite = Value of Row 5

$TrackingID = Value of Row 2

$TrackingUID = Value of Row 3

$QualType = BCID 6333 “mpMRI Assessment Type”

$QualValue = BCID 6335 “mpMRI Assessment Value”

8

>

CONTAINS

INCLUDE

DTID 1411 “Volumetric ROI Measurements and Qualitative Evaluations”

1-n

MC

IF Rows 7, 9 are absent

$Measurement = DT (118565006, SCT, "Volume")

$Units = BCID 7462 “Volume Measurement Unit”

$FindingType = Value of Row 4

$TargetSite = Value of Row 5

$TrackingID = Value of Row 2

$TrackingUID = Value of Row 3

$QualType = BCID 6333 “mpMRI Assessment Type”

$QualValue = BCID 6335 “mpMRI Assessment Value”

9

>

CONTAINS

INCLUDE

DTID 1501 “Measurement and Qualitative Evaluation Group”

1-n

MC

IF Rows 7, 8 are absent

$Measurement = BCID 218 “Quantitative Image Feature”

$Units = BCID 7181 “Abstract Multi-dimensional Image Model Component Unit”

$FindingType = Value of Row 4

$TargetSite = Value of Row 5

$TrackingID = Value of Row 2

$TrackingUID = Value of Row 3

$QualType = BCID 6333 “mpMRI Assessment Type”

$QualValue = BCID 6335 “mpMRI Assessment Value”

$ImagePurpose = EV (260753009, SCT, "Source")

10

>

CONTAINS

CONTAINER

BCID 6300 “Prostate Anatomy”

1-n

U

12

>>

CONTAINS

TEXT

BCID 6333 “mpMRI Assessment Type”

1-n

U

13

>>

CONTAINS

CODE

BCID 6333 “mpMRI Assessment Type”

1-n

U

BCID 6335 “mpMRI Assessment Value”


Content Item Descriptions

Rows 10, 11, 12

These rows are used to describe non-localized findings within the prostate gland (i.e., those that can be described by referring to a specific anatomic part of the gland, but which are not annotated on the image).

DICOM PS3.16 2026a - Content Mapping Resource