DICOM PS3.16 2025d - Content Mapping Resource

TID 4302 Prostate Imaging Findings

Summary of the findings identified in the process of review of prostate imaging study. The container shall include measurement of the prostate gland and at least one localized finding.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 4302. Prostate Imaging Findings

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130553, DCM, "Prostate Imaging Findings")

1

M

2

>

CONTAINS

INCLUDE

DTID 4303 “Overall Prostate Finding”

1

M

3

>

CONTAINS

INCLUDE

DTID 4304 “Localized Prostate Finding”

1-n

M

4

>

CONTAINS

INCLUDE

DTID 4305 “Extra-prostatic Finding”

1-n

U

5

>

CONTAINS

CODE

EV (111005, DCM, "Assessment Category")

1-n

MC

XOR Row 6.

At least one of Rows 5,6,7 shall be present.

DCID 6324 “Prostate MRI Assessment”

6

>

CONTAINS

CODE

EV (RID50294, RADLEX, "PI-RADS Overall Assessment Category")

1-n

MC

XOR Row 5.

At least one of Rows 5,6,7 shall be present.

DCID 6325 “Overall Assessment from PI-RADS®”

7

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

M


Content Item Descriptions

Rows 5, 6

Row 5 can be used to assign an assessment category using an assessment other than PI-RADS®. If PI-RADS® is used for the assessment, Row 6 should be used.

DICOM PS3.16 2025d - Content Mapping Resource