DICOM PS3.16 2026b - Content Mapping Resource

TID 4206 Breast Imaging Report Finding Section

This Template defines a supplementary data section for the Findings of the Breast Imaging Report. It is included from TID 4208 “Breast Imaging Report Supplementary Data”.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 4206. Breast Imaging Report Finding Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

U

3

>

CONTAINS

INCLUDE

DTID 4201 “Breast Imaging Procedure Reported”

1

M

4

>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1-n

M

DCID 6054 “Breast Imaging Finding”

5

>>

HAS CONCEPT MOD

CODE

EV (111405, DCM, "Implant type")

1-n

UC

May be present if value of Row 4 = (40388003, SCT, "Implant")

DCID 6059 “Breast Implant Type”

5b

>>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

U

DCID 6022 “Side”

6

>>

HAS PROPERTIES

INCLUDE

DTID 4203 “Breast Imaging Assessment”

1

U

7

>>

HAS PROPERTIES

CODE

EV (111014, DCM, "Clockface or region")

1

U

DCID 6018 “Clockface Location or Region”

8

>>

HAS PROPERTIES

CODE

EV (111048, DCM, "Quadrant location")

1

U

DCID 6020 “Quadrant Location”

9

>>

HAS PROPERTIES

INCLUDE

DTID 1400 “Linear Measurement”

1-n

U

10

>>

HAS PROPERTIES

INCLUDE

DTID 1401 “Area Measurement”

1-n

U

11

>>

HAS PROPERTIES

INCLUDE

DTID 1402 “Volume Measurement”

1-n

U

12

>>

HAS PROPERTIES

CODE

EV (111020, DCM, "Depth")

1

U

DCID 6024 “Depth”

13

>>

HAS PROPERTIES

CODE

EV (111035, DCM, "Lesion Density")

1

U

DCID 6008 “Density Modifier”

14

>>

HAS PROPERTIES

CODE

EV (107644003, SCT, "Shape")

1-n

U

DCID 6004 “Mammography Shape Characteristic”

15

>>

HAS PROPERTIES

CODE

EV (112233002, SCT, "Margin")

1-n

U

DCID 6006 “Mammography Margin Characteristic”

16

>>

HAS PROPERTIES

CODE

EV (111009, DCM, "Calcification Type")

1-n

U

DCID 6010 “Mammography Calcification Type”

17

>>

HAS PROPERTIES

CODE

EV (111008, DCM, "Calcification Distribution")

1

U

DCID 6012 “Calcification Distribution Modifier”

18

>>

HAS PROPERTIES

NUM

EV (111038, DCM, "Number of calcifications")

1

U

UNITS = EV ({calcifications}, UCUM, "calcifications")

Value = Integer 1 - n

19

>>

HAS PROPERTIES

CODE

EV (111407, DCM, "Implant finding")

1-n

U

DCID 6072 “Breast Implant Finding”

20

>>

HAS PROPERTIES

CODE

EV (246090004, SCT, "Associated Finding")

1-n

U

DCID 6056 “Associated Finding for Breast”

21

>>

HAS PROPERTIES

NUM

EV (111406, DCM, "Number of similar findings")

1

U

UNITS = EV ({findings}, UCUM, "findings")

Value = Integer 2 - n

22

>>

HAS PROPERTIES

CODE

EV (129720009, SCT, "Change since last mammogram")

1-n

U

DCID 6002 “Change Since Last Mammogram or Prior Surgery”

23

>>

HAS PROPERTIES

CODE

EV (111354, DCM, "Orientation")

1

U

DCID 6152 “Orientation”

24

>>

HAS PROPERTIES

CODE

EV (111357, DCM, "Lesion boundary")

1

U

DCID 6153 “Lesion Boundary”

25

>>

HAS PROPERTIES

CODE

EV (111360, DCM, "Echo pattern")

1

U

DCID 6154 “Echo Pattern”

26

>>

HAS PROPERTIES

CODE

EV (111366, DCM, "Posterior acoustic features")

1

U

DCID 6155 “Posterior Acoustic Feature”

27

>>

HAS PROPERTIES

CODE

EV (111371, DCM, "Identifiable effect on surrounding tissues")

1

U

DCID 6015 “Single Image Finding from BI-RADS®”

28

>>

HAS PROPERTIES

CODE

EV (111372, DCM, "Vascularity")

1

U

DCID 6157 “Vascularity”

29

>>

HAS PROPERTIES

CODE

EV (111380, DCM, "Correlation to Other Findings")

1-n

U

DCID 6158 “Correlation to Other Finding”

30

>>

INFERRED FROM

INCLUDE

DTID 350 “References to Supporting Evidence”

1

U


Content Item Descriptions

Row 5b

The laterality of the finding may be different from the laterality of the procedure specified in Row 3, if the latter is undefined or bilateral. If this Content Item is absent, the laterality of the procedure is assumed to apply. The laterality of the finding may be bilateral.

Row 15

In previous releases of the Standard, (111037, DCM, "Margins") was used instead of (112233002, SCT, "Margin"). See PS3.16-2025c.