DICOM PS3.16 2026c - Content Mapping Resource

TID 4122 CAD Common Image Properties Entry

Each instance of the CAD Common Image Properties Entry Template contains selected attributes for a set of parallel contiguous equally spaced slices (with identical properties) from which CAD findings are derived.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 4122. CAD Common Image Properties Entry

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (112224, DCM, "Image Set Properties")

1

M

2

>

CONTAINS

UIDREF

EV (112227, DCM, "Frame of Reference UID")

1

M

3

>

CONTAINS

UIDREF

EV (110180, DCM, "Study Instance UID")

1

M

4

>

CONTAINS

DATE

EV (111060, DCM, "Study Date")

1

M

Shall be taken from Study Date (0008,0020) in the Image Instances.

5

>

CONTAINS

TIME

EV (111061, DCM, "Study Time")

1

M

Shall be taken from Study Time (0008,0030) in the Image Instances.

6

>

CONTAINS

CODE

EV (121139, DCM, "Modality")

1

M

Shall be taken from Modality (0008,0060) in the Image Instances.

7

>

CONTAINS

NUM

EV (111026, DCM, "Horizontal Pixel Spacing")

1

M

Shall be taken from value 2 of Pixel Spacing (0028,0030) in the Image Instances.

UNITS = EV (mm/{pixel}, UCUM, "millimeters per pixel")

8

>

CONTAINS

NUM

EV (111066, DCM, "Vertical Pixel Spacing")

1

M

Shall be taken from value 1 of Pixel Spacing (0028,0030) in the Image Instances.

UNITS = EV (mm/{pixel}, UCUM, "millimeters per pixel")

9

>

CONTAINS

NUM

EV (112225, DCM, "Slice Thickness")

1

M

Shall be taken from Slice Thickness (0018,0050) in the Image Instances.

UNITS = EV (mm, UCUM, "millimeter")

10

>

CONTAINS

NUM

EV (112226, DCM, "Spacing between slices")

1

M

Shall be computed from the Image Position (Patient) (0020,0032) projected onto the normal to the Image Orientation (Patient) (0020,0037); may or may not be the same as the Spacing Between Slices (0018,0088) if present.

UNITS = EV (mm, UCUM, "millimeter")

11

>

CONTAINS

CODE

EV (112228, DCM, "Recumbent Patient Position with respect to gravity")

1

MC

Required if Patient Position (0018,5100) is present in the image instances and has a value.

Shall be derived from Patient Position (0018,5100) in the Image Instances.

DCID 6206 “Recumbent Patient Orientation for Colon”


DICOM PS3.16 2026c - Content Mapping Resource