DICOM PS3.16 2026c - Content Mapping Resource

TID 4014 CAD Image Quality

This Template provides the detail specific to image quality. It allows the encoding of descriptors of image quality (e.g., CID 6041 “Mammography Image Quality Finding”) for a given image or region of an image. For instance, images with partial motion blur can be identified with the region noted.

Table TID 4014. Parameters

Parameter Name

Parameter Usage

$QualityFinding

Coded term or Context Group for Quality Finding

$QualityStandard

Coded term or Context Group for Quality Control Standard


Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4014. CAD Image Quality

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (111052, DCM, "Quality Finding")

1

M

$QualityFinding

2

>

HAS PROPERTIES

CODE

EV (111050, DCM, "Quality Assessment")

1

U

DCID 6044 “Image Quality Assessment Type”

3

>

HAS PROPERTIES

CODE

EV (111051, DCM, "Quality Control Standard")

1

UC

Shall be present if Row 2 is present.

$QualityStandard

4

>

HAS PROPERTIES

NUM

EV (111029, DCM, "Image Quality Rating")

1

U

UNITS = EV (%, UCUM, "Percent")

Value = 0 - 100


Content Item Descriptions

Row 4

A numeric value in the range 0 to 100, inclusive, where 0 is worst quality and 100 is best quality.

DICOM PS3.16 2026c - Content Mapping Resource