DICOM PS3.16 2026d - Content Mapping Resource

TID 4011 Mammography CAD Density

This Template provides the detail specific to a density.

Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4011. Mammography CAD Density

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (111035, DCM, "Lesion Density")

1

MC

At least one of Rows 1, 2, 3, 4, 5 shall be present

DCID 6008 “Density Modifier”

2

CODE

EV (107644003, SCT, "Shape")

1

MC

At least one of Rows 1, 2, 3, 4, 5 shall be present

DCID 6004 “Mammography Shape Characteristic”

3

CODE

EV (112233002, SCT, "Margin")

1-n

MC

At least one of Rows 1, 2, 3, 4, 5 shall be present

DCID 6006 “Mammography Margin Characteristic”

4

INCLUDE

DTID 1400 “Linear Measurement”

1-n

MC

At least one of Rows 1, 2, 3, 4, 5 shall be present

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1400 “Linear Measurement” shall be used.

5

INCLUDE

DTID 1401 “Area Measurement”

1-n

MC

At least one of Rows 1, 2, 3, 4, 5 shall be present

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1401 “Area Measurement” shall be used.

6

INCLUDE

DTID 1402 “Volume Measurement”

1-n

U

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1402 “Volume Measurement” shall be used.


Content Item Descriptions

Row 3

In previous releases of the Standard, (111037, DCM, "Margins") was used instead of (112233002, SCT, "Margin"). See PS3.16-2025c.