DICOM PS3.16 2025e - Content Mapping Resource

TID 4002 Mammography CAD Impression/Recommendation Body

The details of an impression and recommendation are expressed in this Template. It is applied to both Mammography CAD Overall Impression/Recommendation (TID 4001 “Mammography CAD Overall Impression/Recommendation”) and Mammography CAD Individual Impression/Recommendation (TID 4003 “Mammography CAD Individual Impression/Recommendation”).

Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4002. Mammography CAD Impression/Recommendation Body

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (111005, DCM, "Assessment Category")

1-n

MC

At least one of Rows 1, 3, 5, 6, 8, 9 shall be present.

DCID 6026 “Mammography Assessment”

2

>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

U

DCID 6022 “Side”

3

CODE

EV (111023, DCM, "Differential Diagnosis/ Impression")

1-n

MC

At least one of Rows 1, 3, 5, 6, 8, 9 shall be present.

DCID 6002 “Change Since Last Mammogram or Prior Surgery”

4

>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

U

DCID 6022 “Side”

5

TEXT

EV (111033, DCM, "Impression Description")

1

MC

At least one of Rows 1, 3, 5, 6, 8, 9 shall be present.

6

CODE

EV (111053, DCM, "Recommended Follow-up")

1-n

MC

At least one of Rows 1, 3, 5, 6, 8, 9 shall be present.

DCID 6028 “Mammography Recommended Follow-up”

7

>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

U

DCID 6022 “Side”

8

NUM

EV (111055, DCM, "Recommended Follow-up Interval")

1

MC

At least one of Rows 1, 3, 5, 6, 8, 9 shall be present. May be present only if (111054, DCM, "Recommended Follow-up Date") is not present.

UNITS = DCID 6046 “Follow-up Interval Unit”

Values = Integer ≥ 0, where 0 = immediate follow-up

9

DATE

EV (111054, DCM, "Recommended Follow-up Date")

1

MC

At least one of Rows 1, 3, 5, 6, 8, 9 shall be present. May be present only if (111055, DCM, "Recommended Follow-up Interval") is not present.

Shall be later than date of study

10

NUM

EV (111013, DCM, "Certainty of impression")

1

UC

May be present only if (111005, DCM, "Assessment Category"), (111023, DCM, "Differential Diagnosis/Impression") or (111033, DCM, "Impression Description") is present.

UNITS = EV (%, UCUM, "Percent")

Values = 0 - 100

11

INCLUDE

DTID 4019 “Algorithm Identification”

1-n

M

12

NUM

DCID 6142 “Calculated Value”

1-n

U

13

>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

U

DCID 6022 “Side”

14

>

HAS CONCEPT MOD

CODE

EV (121401, DCM, "Derivation")

1

M

DCID 6140 “Calculation Method”

15

>

INFERRED FROM

TEXT

EV (112034, DCM, "Calculation Description")

1

U


Content Item Descriptions

Row 5

Free-form text describing the overall or an individual impression

Row 10

The certainty that the device populating the Mammography CAD SR report places on this impression, where 0 equals no certainty and 100 equals certainty.

DICOM PS3.16 2025e - Content Mapping Resource