DICOM PS3.16 2026d - Content Mapping Resource

TID 3929 Cardiovascular Analysis Observation Context

Defines the observation context for cardiovascular Functional Analysis

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3929. Cardiovascular Analysis Observation Context

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

HAS OBS CONTEXT

NUM

EV (8867-4, LN, "Heart Rate")

1

U

UNITS = DT ({H.B.}/min, UCUM, "BPM")

2

HAS OBS CONTEXT

CODE

EV (8884-9, LN, "Cardiac Rhythm")

1

U

DCID 3826 “Heart Rhythm”

3

HAS OBS CONTEXT

NUM

EV (271649006, SCT, "Systolic Blood Pressure")

1

U

UNITS = DT (mm[Hg], UCUM, "mmHg")

4

HAS OBS CONTEXT

NUM

EV (271650006, SCT, "Diastolic Blood Pressure")

1

U

UNITS = DT (mm[Hg], UCUM, "mmHg")

5

HAS OBS CONTEXT

CODE

EV (364062005, SCT, "Respiration Observable")

1

U

DCID 3823 “Respiratory Status”

6

HAS ACQ CONTEXT

INCLUDE

DTID 3106 “Drugs/Contrast Administered”

1-n

U


DICOM PS3.16 2026d - Content Mapping Resource