DICOM PS3.16 2026c - Content Mapping Resource

TID 3908 Vascular Lesion

Specifies properties and the features of a vascular lesion detected during the analysis. In addition it is possible to reference or include growing of lesions over time by adding references to previous reports or by adding previous examination results.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3908. Vascular Lesion

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

CONTAINER

EV (300577008, SCT, "Lesion Finding")

1

M

2

>

CONTAINS

TEXT

EV (121151, DCM, "Lesion Identifier")

1

M

3

>

CONTAINS

INCLUDE

DTID 3909 “Best Illustration of Findings”

1-n

U

4

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1-n

U

5

>

CONTAINS

NUM

EV (122337, DCM, "Relative Position")

1

U

UNITS = EV (mm, UCUM, "mm")

6

>>

HAS CONCEPT MOD

CODE

EV (122340, DCM, "Fiducial Feature")

1

M

DCID 3837 “Fiducial Feature”

7

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (397413000, SCT, "Vessel Lumen Diameter")

$Derivation = DCID 3838 “Diameter Derivation”

$FindingSite = DCID 3486 “Vascular Measurement Site”

$Units = DT (mm, UCUM, "mm")

8

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (397415007, SCT, "Vessel Lumen Cross-Sectional Area")

$Derivation = DCID 3838 “Diameter Derivation”

$FindingSite = DCID 3486 “Vascular Measurement Site”

$Units = DT (mm2, UCUM, "mm2")

9

>

CONTAINS

CODE

EV (116676008, SCT, "Associated Morphology")

1-n

M

DCID 3810 “Vascular Morphology”

10

>>

INCLUDE

DTID 3909 “Best Illustration of Findings”

1-n

U

11

>>

HAS PROPERTIES

TEXT

EV (121106, DCM, "Comment")

1-n

U

12

>>

INCLUDE

DTID 3911 “Plaque Properties”

1

MC

IFF value of Row 9 = (1522000, SCT, "Plaque")

13

>>

INCLUDE

DTID 3912 “Stenosis Properties”

1

MC

IFF value of Row 9 = (415582006, SCT, "Stenosis")

14

>>

INCLUDE

DTID 3913 “Aneurysm Properties”

1

MC

IFF value of Row 9 = (85659009, SCT, "Aneurysm")

15

>>

INCLUDE

DTID 3914 “Arterial Dissection Properties”

1

MC

IFF value of Row 9 = (710864009, SCT, "Arterial Dissection")

16

>>

HAS PROPERTIES

CODE

EV (116676008, SCT, "Associated Morphology")

1

MC

IFF value of Row 9 = (107671003, SCT, "Vascular Sclerosis")

DCID 3817 “Vascular Sclerosis Type”

17

>>

INCLUDE

DTID 3915 “Vascular Occlusion Properties”

1

MC

IFF value of Row 9 = (26036001, SCT, "Occlusion")

18

>>

INCLUDE

DTID 3916 “Stent Properties”

1

MC

IFF value of Row 9 = (65818007, SCT, "Stent")


Content Item Descriptions

Row 5

A positive value indicates a distance in the direction of flow within the vessel

For example: An aneurysm with relative position -4 mm from the renal arteries would begin superior to the renal arteries.

Rows 7, 8

These rows are associated with the position of the most significant effect of the lesion, i.e., maximum diameter of aneurysm or the minimum diameter of stenosis