DICOM PS3.16 2026c - Content Mapping Resource

TID 3905 Calcium Scoring Results

Contains the calcium scoring results related to plaque findings, vessels or the whole body.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3905. Calcium Scoring Results

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

EV (122603, DCM, "Calcium Scoring Analysis")

3

>

CONTAINS

NUM

EV (122657, DCM, "Agatston Score Threshold")

1

U

UNITS = DT ([hnsf'U], UCUM, "Hounsfield unit")

4

>

CONTAINS

NUM

EV (122658, DCM, "Calcium Mass Threshold")

1

U

UNITS = DT (mg/cm3, UCUM, "mg/cm3")

5

>

CONTAINS

NUM

EV (122659, DCM, "Calcium Scoring Calibration")

1

U

UNITS = DT (mg/[hnsf'U].cm3, UCUM, "mg/[hnsf'U].cm3")

6

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (450360000, SCT, "Coronary artery calcium score")

$Method = EV (112055, DCM, "Agatston Scoring Method")

$Units = DT (1, UCUM, "no units")

7

>

CONTAINS

NUM

EV (122660, DCM, "Calcium Volume")

1

U

UNITS = EV (mm3, UCUM, "mm3")

8

>

CONTAINS

NUM

EV (122661, DCM, "Calcium Mass")

1

U

UNITS = EV (mg, UCUM, "mg")

9

>

CONTAINS

NUM

EV (246206008, SCT, "Number of Lesions")

1

U

UNITS = DT ({lesions}, UCUM, "lesions")

10

>

CONTAINS

INCLUDE

DTID 3909 “Best Illustration of Findings”

1-n

U


Content Item Descriptions

Row 2

Previously, the relationship was specified as CONTAINS rather than HAS CONCEPT MOD, which was inconsistent with other uses in the family of cardiovascular analysis results and similar uses in other families of templates. The usage of CONTAINS for this Row has been retired. See PS3.16-2025c.