DICOM PS3.16 2026a - Content Mapping Resource

TID 3815 Right Ventriculography Findings

The information contained here about right ventricle is relatively qualitative in nature. It is a high-level summary of the more detailed information that may be contained in an optional Quantitative Ventricular Analysis report.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3815. Right Ventriculography Findings

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure reported")

1

M

DT (265483003, SCT, "Right Ventriculography")

3

>

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (10231-9, LN, "RV Ejection Fraction")

$Units = EV (%, UCUM, "%")

$Method = DCID 3748 “Angiographic EF Testing Method”

$Derivation = DCID 3745 “Calculation Method”

4

>

CONTAINS

CODE

EV (250964004, SCT, "Right Ventricular Cavity Size")

1

U

DCID 3705 “Chamber Size”

5

>

CONTAINS

CODE

EV (250949004, SCT, "Right Ventricular Contractility")

1

U

DCID 3706 “Overall Contractility”

6

>

CONTAINS

INCLUDE

DTID 3816 “Ventricular Assessment”

1

U


DICOM PS3.16 2026a - Content Mapping Resource