1
|
|
|
CONTAINER
|
EV (122178, DCM, "Lesion Intervention Information")
|
1
|
M
|
|
|
|
2
|
>
|
CONTAINS
|
INCLUDE
|
DTID 3105 “Lesion Identification and Properties”
|
1
|
M
|
|
|
|
3
|
>
|
CONTAINS
|
CODE
|
EV (108, NDR [2.0b], "Previous Dilation")
|
1
|
U
|
|
DCID 3750 “Previously Dilated Lesion”
|
|
4
|
>
|
CONTAINS
|
CODE
|
EV (103, NDR [2.0b], "Guidewire crossing lesion")
|
1
|
U
|
|
DCID 3752 “Guidewire Crossing”
|
|
5
|
>
|
CONTAINS
|
CODE
|
EV (116682006, SCT, "Uses Equipment")
|
1-n
|
M
|
|
DCID 3411 “Intracoronary Device”
|
|
6
|
>>
|
HAS CONCEPT MOD
|
CODE
|
EV (122111, DCM, "Primary Intervention Device")
|
1
|
MC
|
IF Device is Primary for this Lesion
|
DCID 230 “Yes-No”
|
|
7
|
>>
|
HAS PROPERTIES
|
TEXT
|
EV (121145, DCM, "Description of Material")
|
1
|
U
|
|
|
|
8
|
>>
|
HAS PROPERTIES
|
NUM
|
DCID 3423 “Numeric Device Characteristic”
|
1-n
|
U
|
|
|
|
9
|
>>
|
HAS PROPERTIES
|
NUM
|
DCID 3425 “Intervention Parameter”
|
1-n
|
U
|
|
|
|
10
|
>>
|
HAS PROPERTIES
|
DATETIME
|
EV (122105, DCM, "DateTime of Intervention")
|
1
|
U
|
|
|
|
11
|
>>
|
HAS PROPERTIES
|
NUM
|
EV (122106, DCM, "Duration of Intervention")
|
1
|
U
|
|
UNITS = EV (s, UCUM, "s")
|
|
12
|
>
|
CONTAINS
|
NUM
|
EV (408715008, SCT, "Lumen Diameter Stenosis")
|
1
|
M
|
|
UNITS = EV (%, UCUM, "%")
|
|
13
|
>>
|
HAS CONCEPT MOD
|
CODE
|
EV (129085009, SCT, "Catheterization Procedure Phase")
|
1
|
M
|
|
EV (128960007, SCT, "Post-intervention Phase")
|
|
14
|
>>
|
HAS CONCEPT MOD
|
CODE
|
EV (121401, DCM, "Derivation")
|
1
|
U
|
|
DCID 3745 “Calculation Method”
|
|
15
|
>>
|
|
INCLUDE
|
DTID 1000 “Quotation”
|
1
|
U
|
|
|
|
16
|
>
|
CONTAINS
|
CODE
|
EV (122110, DCM, "Post-Intervention TIMI Flow")
|
1
|
U
|
|
DCID 3713 “TIMI Flow Characteristic”
|
|
17
|
>
|
CONTAINS
|
CODE
|
EV (115, NCDR [2.0b], "Dissection in segment observed")
|
1
|
U
|
|
DCID 230 “Yes-No”
|
|
18
|
>
|
CONTAINS
|
CODE
|
EV (116, NCDR [2.0b], "Acute closure observed")
|
1
|
U
|
|
DCID 230 “Yes-No”
|
|
19
|
>
|
CONTAINS
|
CODE
|
EV (117, NCDR [2.0b], "Acute closure re-opened")
|
1
|
UC
|
IFF Row 18 value is <yes>
|
DCID 230 “Yes-No”
|
|
20
|
>
|
CONTAINS
|
CODE
|
EV (118, NCDR [2.0b], "Perforation occurred")
|
1
|
U
|
|
DCID 230 “Yes-No”
|
|
21
|
>
|
CONTAINS
|
IMAGE
|
DT (121080, DCM, "Best illustration of finding")
|
1
|
U
|
|
|
|
22
|
>
|
CONTAINS
|
TEXT
|
DT (122177, DCM, "Procedure Result")
|
1
|
U
|
|
|