DICOM PS3.16 2026a - Content Mapping Resource

TID 3708 ECG Waveform Information

The ECG Waveform Information Template provides reference to, and important parameters of, the analyzed waveform.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3708. ECG Waveform Information

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (55111-9, LN, "Current Procedure Descriptions")

1

M

2

>

CONTAINS

WAVEFORM

EV (121112, DCM, "Source of Measurement")

1

U

3

>

CONTAINS

CODE

EV (10:11345, MDC, "Lead System")

1

U

DCID 3263 “Electrode Placement Value”

4

>

CONTAINS

TEXT

EV (122142, DCM, "Acquisition Device Type")

1

U

5

>

CONTAINS

TEXT

EV (121122, DCM, "Equipment Identification")

1

U

6

>

CONTAINS

INCLUDE

DTID 1003 “Person Observer Identifying Attributes”

1

U

Person performing the ECG acquisition

7

>

CONTAINS

TEXT

EV (121121, DCM, "Room Identification")

1

U

8

>

CONTAINS

DATETIME

EV (122146, DCM, "Procedure DateTime")

1

M

9

>

CONTAINS

NUM

DCID 3690 “ECG Control Numeric Variable”

1-n

U

10

>

CONTAINS

TEXT

DCID 3691 “ECG Control Text Variable”

1-n

U

11

>

CONTAINS

INCLUDE

DTID 4019 “Algorithm Identification”

1

U


DICOM PS3.16 2026a - Content Mapping Resource