DICOM PS3.16 2026c - Content Mapping Resource

TID 3516 Blood Lab Measurements

The Blood Lab Measurements Template provides for the recording of measurements made on blood samples obtained during a catheterization procedure. The type and anatomic source of the blood is recorded as acquisition context. The results from the blood chemistry measurement system are quoted; the measurement names may be pre-coordinated with the type or source of the blood, or generic measurement names may be reported. In the latter case, the full measurement concept name may be effectively post-coordinated using the recorded acquisition context.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3516. Blood Lab Measurements

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122125, DCM, "Blood lab measurements")

1

M

2

>

HAS ACQ CONTEXT

CODE

EV (371439000, SCT, "Specimen Type")

1

M

DCID 3520 “Blood Source Type”

3

>

HAS ACQ CONTEXT

CODE

EV (363704007, SCT, "Procedure site")

1

M

BCID 3630 “Cardiovascular Anatomic Location”

4

>

INCLUDE

DTID 1000 “Quotation”

1

U

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (718-7, LN, "Hemoglobin")

$Units = DT (g/dl, UCUM, "g/dl")

6

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3524 “Blood Gas Pressure”

$Units = DCID 3500 “Pressure Unit”

7

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3525 “Blood Gas Content”

$Units = DT (ml/dl, UCUM, "ml/dl")

8

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3528 “Blood pH”

$Units = EV ([pH], UCUM, "pH")

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3526 “Blood Gas Saturation”

$Units = EV (%, UCUM, "%")

10

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3527 “Blood Base Excess”

$Units = DT (meq/dl, UCUM, "meq/dl")

11

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122183, DCM, "Blood temperature")

$Units = EV (Cel, UCUM, "C")


DICOM PS3.16 2026c - Content Mapping Resource