DICOM PS3.16 2026c - Content Mapping Resource

TID 3515 Cardiac Output Measurement by Indicator Dilution

The Cardiac Output Measurement by Indicator Dilution Template consists of a CONTAINER containing the measurement and significant parameters of the technical method. If the measurement is based on a DICOM Hemodynamic Waveform SOP Instance, that object may also be referenced.

Note

Fick Cardiac Output is encoded in TID 3560 “Derived Hemodynamic Measurements”.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3515. Cardiac Output Measurement By Indicator Dilution

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (117610000, SCT, "Cardiac Output measurement")

1

M

2

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (8737-9, LN, "Cardiac Output by Indicator Dilution")

$Units = EV (l/min, UCUM, "l/min")

$Method = DCID 3628 “Cardiac Output Method”

$WavePurpose = DT (121112, DCM, "Source of Measurement")

3

>

HAS ACQ CONTEXT

NUM

EV (122319, DCM, "Catheter Size")

1

MC

IFF Row 2 indicates a thermal method

UNITS = DCID 3510 “Catheter Size Unit”

4

>

HAS ACQ CONTEXT

NUM

EV (122320, DCM, "Injectate Temperature")

1

MC

IFF Row 2 indicates a thermal method

UNITS = EV (Cel, UCUM, "C")

5

>

HAS ACQ CONTEXT

NUM

EV (122321, DCM, "Injectate Volume")

1

M

UNITS = DT (ml, UCUM, "ml")

6

>

HAS ACQ CONTEXT

NUM

EV (122322, DCM, "Calibration Factor")

1

M

UNITS = DT (1, UCUM, "no units")


DICOM PS3.16 2026c - Content Mapping Resource