DICOM PS3.16 2026c - Content Mapping Resource

TID 3509 Blood Velocity Measurement

The Blood Velocity Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic site, and containing primary (not derived) velocity measurements, e.g., from a Doppler catheter. Derived velocity measurements may be recorded using TID 3560 “Derived Hemodynamic Measurements”.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3509. Blood Velocity Measurement

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122124, DCM, "Blood velocity measurements")

1

M

2

>

INCLUDE

DTID 3530 “Hemodynamic Acquisition Context”

1

M

$LocationName = EV (363704007, SCT, "Procedure site")

$LocationValue = BCID 3606 “Arterial Source Location”

3

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

M

$Measurement = DCID 3612 “Blood Velocity Measurement”

$Units = EV (mm/s, UCUM, "mm/s")

4

>

CONTAINS

INCLUDE

DTID 3550 “Pressure Waveform Measurements”

1-n

U