DICOM PS3.16 2026c - Content Mapping Resource

TID 3303 Stress Test Phase Data

The Stress Test Phase Data Template provides a structure for measurements acquired during a single procedure phase.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3303. Stress Test Phase Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (59776-5, LN, "Findings")

1

M

2

>

HAS ACQ CONTEXT

CODE

EV (128954007, SCT, "Procedure phase")

1

MC

XOR Row 3

BCID 3207 “Stress Test Procedure Phase”

3

>

HAS ACQ CONTEXT

CODE

EV (128954007, SCT, "Procedure phase")

1

MC

IFF Nuclear Imaging; XOR Row 2

DCID 3101 “Cardiac Procedural State Value”

4

>

HAS ACQ CONTEXT

TEXT

EV (128954007, SCT, "Procedure phase")

1

U

5

>

CONTAINS

INCLUDE

DTID 3301 “Stress Test Procedure Description”

1

MC

IFF protocol changed from initial specification

6

>

HAS ACQ CONTEXT

NUM

EV (109055, DCM, "Protocol Stage")

1

U

UNITS = DT ({stage}, UCUM, "stage")

7

>

CONTAINS

INCLUDE

DTID 3304 “Stress Test Measurement Group”

1-n

U

8

>

CONTAINS

INCLUDE

DTID 3307 “NM/PET Perfusion Measurement Group”

1

MC

IFF Nuclear or PET Imaging

9

>

CONTAINS

INCLUDE

DTID 3309 “Stress Echo Measurement Group”

1

UC

IFF Echocardiography Imaging


Content Item Descriptions

Row 1

The Container shall have a specific Content Item Observation DateTime (0040,A032) Attribute to indicate the time at which the phase began.

DICOM PS3.16 2026c - Content Mapping Resource