DICOM PS3.16 2025e - Content Mapping Resource

TID 3253 IVUS Measurements

The IVUS measurements Template groups together simple distance, area and angle measurements, along with derived measurements that made during an IVUS procedure. Refer to the " ACC Clinical Expert Consensus Document on Standards for Acquisition, measurement and Reporting of Intravascular Ultrasound Studies (IVUS) "for more information.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3253. IVUS Measurements

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3481 “IVUS Distance Measurement”

$Units = EV (mm, UCUM, "mm")

$Derivation = DCID 3488 “Min/Max/Mean”

$TargetSite = BCID 3486 “Vascular Measurement Site”

2

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3482 “IVUS Area Measurement”

$Units = EV (mm2, UCUM, "mm2")

$Derivation = DCID 3488 “Min/Max/Mean”

$TargetSite = BCID 3486 “Vascular Measurement Site”

3

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3483 “IVUS Longitudinal Measurement”

$Units = EV (mm, UCUM, "mm")

4

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (122355, DCM, "Arc of Calcium")

$Units = EV (deg, UCUM, "degrees")

$TargetSite = BCID 3486 “Vascular Measurement Site”

5

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (408714007, SCT, "Lumen Area Stenosis")

$Units = EV (%, UCUM, "%")

6

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122354, DCM, "Plaque Burden")

$Units = EV (%, UCUM, "%")

$TargetSite = BCID 3486 “Vascular Measurement Site”

7

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3484 “IVUS Index/Ratio”

$Units = EV ({ratio}, UCUM, "ratio")

$TargetSite = BCID 3486 “Vascular Measurement Site”

8

INCLUDE

DTID 3255 “IVUS Volume Measurement”

1-n

U

9

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122339, DCM, "Stent Volume Obstruction")

$Units = EV (%, UCUM, "%")


DICOM PS3.16 2025e - Content Mapping Resource