DICOM PS3.16 2026d - Content Mapping Resource

TID 3105 Lesion Identification and Properties

The Lesion Identification and Properties Template allows recording the identification of each lesion addressed in a procedure. The lesion identifier may be used to relate diagnostic or therapeutic actions with their target lesion (see Row 4 in TID 3010 “Log Entry Qualifiers”). This Content Item may include the initial visually estimated measurements of stenosis or TIMI flow; measured values from a quantitative measurement report may be referenced indirectly (through TID 3103 “Referenced Object”), or by quotation (TID 3109 “Measurements”). Subsequent (e.g., post-intervention) stenosis measurements may be encoded using TID 3109 “Measurements”, with the Lesion Identifier conveyed through its subsidiary TID 3010 “Log Entry Qualifiers” Template.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3105. Lesion Identification and Properties

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

TEXT

EV (121151, DCM, "Lesion Identifier")

1

M

Up to 3 numeric characters

2

>

HAS PROPERTIES

CODE

EV (363698007, SCT, "Finding Site")

1

M

DCID 3604 “Arterial Lesion Location”

3

>>

HAS CONCEPT MOD

CODE

EV (106233006, SCT, "Topographical modifier")

1

U

DCID 3019 “Cardiovascular Anatomic Structure Modifier”

4

>

HAS PROPERTIES

CODE

EV (121153, DCM, "Lesion Risk")

1

U

DCID 3418 “Lesion Risk”

5

>

HAS PROPERTIES

NUM

EV (408715008, SCT, "Lumen Diameter Stenosis")

1

U

UNITS = EV (%, UCUM, "%")

6

>>

HAS CONCEPT MOD

CODE

EV (129085009, SCT, "Catheterization Procedure Phase")

1

M

EV (128955008, SCT, "Baseline Phase")

7

>>

HAS CONCEPT MOD

CODE

EV (121401, DCM, "Derivation")

1

U

DCID 3745 “Calculation Method”

8

>

HAS PROPERTIES

CODE

EV (122109, DCM, "Baseline TIMI Flow")

1

UC

IFF Row 2 specifies a coronary artery

DCID 3713 “TIMI Flow Characteristic”

9

>

HAS PROPERTIES

CODE

EV (122131, DCM, "Degree of Thrombus")

1

U

DCID 3714 “Thrombus”

10

>

HAS PROPERTIES

CODE

EV (129737002, SCT, "Lesion Margin Characteristics")

1

U

DCID 3715 “Lesion Margin”

11

>

HAS PROPERTIES

CODE

EV (122134, DCM, "Vessel Morphology")

1-n

U

DCID 3712 “Vessel Descriptor”

12

>

HAS PROPERTIES

CODE

EV (122132, DCM, "Severity of Calcification")

1

U

DCID 3716 “Severity”

13

>

HAS PROPERTIES

IMAGE

DT (121080, DCM, "Best illustration of finding")

1

U

14

>

INCLUDE

DTID 3010 “Log Entry Qualifiers”

1

U


Content Item Descriptions

Row 1

Lesion Identifier is specified as a numeric text string in order to facilitate transcoding to DICOM Attribute (0018,3105) Lesion Number and to formats for outcomes registries, such as the ACC National Cardiovascular Data Registry™.

DICOM PS3.16 2026d - Content Mapping Resource