Table TID 2123. Circumpapillary Retinal Nerve Fiber Layer Key Measurements
|
EV (131242, DCM, "Circumpapillary Retinal Nerve Fiber Layer Key Measurements") |
||||||||
|
$Method = DCID 4282 “Retinal Sector Methods” $Measurement = EV (131274, DCM, "Retinal ROI width") $OptMeasurement = DCID 4283 “RNFL Sector Measurements” |
||||||||
|
$Method = EV (131308, DCM, "RNFL Clockface Method") $Measurement = DCID 4284 “RNFL Clockface Measurements” |
||||||||
|
$Measurement = EV (131273, DCM, "Retinal nerve fiber layer symmetry") |
|
For example, TID 2120 may be invoked once for each eye, or multiple times for a trend series for a single eye. Applications also use various approaches to identifying different sectors of the retina when measuring the RNFL thickness. The sector definition used is specified by a concept from CID 4282 “Retinal Sector Methods”. Measurements that match the sector names defined by the method may be selected from CID 4283 “RNFL Sector Measurements” to be included in the key measurements. Receiving applications are cautioned about comparing or aggregating measurements from measurement groups with different ROI areas or sector methods. |