This Template is a proper subset of TID 1501, with some optional extensions, and may be processed by a receiving application in the same way. The parameters used in this Template are identical to those parameters as used in TID 1501. However, this Template is specialized for ophthalmology (finding site "Eye") , and makes mandatory a Content Item for each measuement or finding concept specified in the invoking Template in the Context Groups invoked for the $Measurement parameter (TID 1501 has no mandatory content).
Table TID 2120. Parameters
Table TID 2120. Ophthalmology Measurements Group
|
Mandatory numeric findings of the measurement group. Each Concept specified in the Value Set Constraints (i.e., as specified in the invoking Template $Measurement parameter) shall be encoded in a NUM Content Item. Thus if there are eight concepts in the specified Context Group, eight NUM Content Items in accordance with TID 300 will be present. Note that the NUM Content Item allows an absent value with an associated reason code per CID 42, e.g., (114007, DCM, "Measurement not attempted"). TID 300 “Measurement” defines an optional capability to specify properties of a measurement via TID 310 “Measurement Properties”. TID 310 supports properties such as normality, statistical properties (through subsidiary TID 311), normal ranges (subsidiary TID 312) , level of significance and more. Normality flags are highly useful and commonly provided by implementations. |
|
|
Optional table of numeric measurements of the measurement group. |
|
|
Reference to the original image(s) , e.g., Ophthalmic Tomography, that provided the data analyzed to produce the measurements in this group. |
|
|
A numeric (row 12) or categorical (row 13) rating of the quality of the source images for the purpose of producing the measurements in this group. |
|
|
May be a reference to a Secondary Capture Image (row 14) or Encapsulated PDF (row 15) rendering of the set of measurements and findings encoded in this Measurement Group, and possibly additional data. |