DICOM PS3.16 2026c - Content Mapping Resource

TID 2100 Macular Grid Thickness and Volume Report

The Macular Grid Thickness and Volume Report is a structured report encoding the macular grid thickness and volume values derived from ophthalmic images, such as ophthalmic OPT images. This may encode measurements of either or both eyes.

The macular grid conveyed by this report is based upon the grid employed by the Early Treatment of Diabetic Retinopathy Study (ETDRS) to measure area and proximity of macular edema to the anatomic center (fovea) of the macula. See ETDRS Report Number 10.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 2100. Macular Grid Thickness and Volume Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (111690, DCM, "Macular Grid Thickness and Volume Report")

1

M

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

M

4

>

CONTAINS

INCLUDE

DTID 2101 “Macular Grid Thickness and Volume Measurement”

1

MC

IF Row 5 is absent.

$Laterality = EV (24028007, SCT, "Right")

5

>

CONTAINS

INCLUDE

DTID 2101 “Macular Grid Thickness and Volume Measurement”

1

MC

IF Row 4 is absent.

$Laterality = EV (7771000, SCT, "Left")


DICOM PS3.16 2026c - Content Mapping Resource