DICOM PS3.16 2026c - Content Mapping Resource

TID 2007 Imaging Procedure Description

Contains information related to the procedure.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 2007. Imaging Procedure Description

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

U

2

CONTAINS

TEXT

EV (123014, DCM, "Target Region")

1

MC

XOR with Row 3

3

CONTAINS

CODE

EV (123014, DCM, "Target Region")

1

MC

XOR with Row 2

DCID 4028 “Craniofacial Anatomic Region”

DCID 4030 “CT, MR and PET Anatomy Imaged”

DCID 4031 “Common Anatomic Region”

4

>

HAS CONCEPT MOD

CODE

EV (122142, DCM, "Acquisition Device Type")

1-n

U

DCID 29 “Acquisition Modality”

5

CONTAINS

TEXT

EV (121065, DCM, "Procedure Description")

1

M

6

CONTAINS

DATE

EV (111060, DCM, "Study Date")

1

M

Shall be equal to the Study Date (0020,0020) in the General Study Module in the images to which this report applies.

7

CONTAINS

TIME

EV (111061, DCM, "Study Time")

1

U

If present, shall be equal to the Study Time (0020,0030) in the General Study Module in the images to which this report applies.

8

CONTAINS

COMPOSITE

EV (113701, DCM, "X-Ray Radiation Dose Report")

1-n

U