DICOM PS3.16 2026a - Content Mapping Resource

TID 10004 Accumulated Fluoroscopy and Acquisition Projection X-Ray Dose

This general Template provides detailed information on projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10004. Accumulated Fluoroscopy and Acquisition Projection X-Ray Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (113726, DCM, "Fluoro Dose Area Product Total")

1

MC

IFF TID 10003 Row 7 value = (44491008, SCT, "Fluoroscopy") for at least one irradiation event

UNITS = EV (Gy.m2, UCUM, "Gy.m2")

2

NUM

EV (113728, DCM, "Fluoro Dose (RP) Total")

1

MC

IFF TID 10003 Row 7 value = (44491008, SCT, "Fluoroscopy") for at least one irradiation event AND any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content").

UNITS = EV (Gy, UCUM, "Gy")

3

NUM

EV (113730, DCM, "Total Fluoro Time")

1

MC

IFF TID 10003 Row 7 value = (44491008, SCT, "Fluoroscopy") for at least one irradiation event.

UNITS = EV (s, UCUM, "s")

4

NUM

EV (113727, DCM, "Acquisition Dose Area Product Total")

1

MC

IF any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content").

UNITS = EV (Gy.m2, UCUM, "Gy.m2")

5

NUM

EV (113729, DCM, "Acquisition Dose (RP) Total")

1

MC

IF any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content").

UNITS = EV (Gy, UCUM, "Gy")

6

NUM

EV (113855, DCM, "Total Acquisition Time")

1

MC

IF any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content").

UNITS = EV (s, UCUM, "s")


Content Item Descriptions

Rows 1, 2, 3

Fluoroscopic component only

Row 3

Total clock time of Fluoroscopy accumulated over the defined scope of accumulation (i.e., the sum of the Irradiation Duration values for accumulated fluoroscopy irradiation events)

Rows 4, 5, 6

Acquisition component only

Row 6

Total clock time of acquisitions accumulated over the defined scope of accumulation (i.e., the sum of the Irradiation Duration values for accumulated acquisition irradiation events)

DICOM PS3.16 2026a - Content Mapping Resource