DICOM PS3.16 2025a - Content Mapping Resource |
---|
This Template contains data that is expected to be available to the X-Ray source component of the equipment.
Table TID 10003B. Irradiation Event X-Ray Source Data
IFF (TID 10001 Row 2 = (113704, DCM, "Projection X-Ray")) AND IF any of the values of TID 10001 Row 18 are not (113858, DCM, "MPPS Content") |
||||||||
IFF TID 10001 Row 2 = (71651007, SCT, "Mammography") |
||||||||
IFF TID 10003 Row 7 value = (44491008, SCT, "Fluoroscopy") |
||||||||
IFF Row 5 value = (113631, DCM, "Pulsed") |
||||||||
IFF Row 5 is not present or Row 5 is present and equals (113631, DCM, "Pulsed") |
||||||||
Required if the irradiating device is not the recording device. |
$DeviceProcedureRole = EV (113859, DCM, "Irradiating Device") |
Dose applied by this irradiation event, relative to defined reference point. |
|
If a precise count of pulses is not available, an estimated number shall be provided, and the Row 8 Concept Modifier shall indicate "Estimated" |
|
Pulse width as measured/recorded by the system, either as a single total value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". |
|
KVP value as measured/recorded by system, either as a single mean value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". |
|
Tube current as measured/recorded by system, either as a single mean value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". |
|
Exposure as measured/recorded by system, either as a single total value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". The Exposure will be affected by the shape of the pulse and other factors, and may not be a simple multiplication of tube current and exposure time. |
|
If one or more Filter(s) were applied during this irradiation event |
|
The device that produced the irradiation in this Irradiation Event. I.e., the X-Ray source. This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10001 “Projection X-Ray Radiation Dose” Row 5, which in turn may be absent if identical to the content in the Enhanced General Equipment Module. |
DICOM PS3.16 2025a - Content Mapping Resource |
---|