Quantitative Ventricular Analysis Report SR IOD Templates 

The Templates that comprise the Quantitative Ventricular Analysis SR are interconnected as in Figure A-3:

Quantitative Ventricular Analysis Report SR IOD Template Structure

Figure A-3. Quantitative Ventricular Analysis Report SR IOD Template Structure


TID 3202 Quantitative Ventricular Analysis Report

The Quantitative Ventricular Analysis Report Template provides a CONTAINER with a structure for reporting the result of the ventricular analysis.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 3202. Quantitative Ventricular Analysis Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122292, DCM, "Quantitative Ventriculography Report")

1

M

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

M

4

>

CONTAINS

INCLUDE

DTID 3602 “Cardiovascular Patient Characteristics”

1

U

5

>

CONTAINS

CONTAINER

EV (122144, DCM, "Quantitative Analysis")

1-n

M

6

>>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

U

7

>>

HAS OBS CONTEXT

TEXT

EV (111001, DCM, "Algorithm Name")

1

M

8

>>

HAS OBS CONTEXT

TEXT

EV (111003, DCM, "Algorithm Version")

1

M

9

>>

HAS OBS CONTEXT

TEXT

EV (122405, DCM, "Algorithm Manufacturer")

1

M

10

>>

CONTAINS

IMAGE

EV (121112, DCM, "Source of Measurement")

1-n

M

11

>>>

HAS CONCEPT MOD

CODE

EV (246092007, SCT, "Cardiac Phase")

1

M

DCID 12233 “Cardiac Phase”

12

>>>

HAS CONCEPT MOD

CODE

EV (111031, DCM, "Image View")

1

MC

If Biplane Analysis

DCID 3466 “Plane Identification”

13

>>

HAS ACQ CONTEXT

INCLUDE

DTID 3205 “Calibration”

1-2

U

VM = 1: Single plane analysis, VM = 2: Biplane analysis

$CalibrationPlane = DCID 3466 “Plane Identification”

14

>>

HAS ACQ CONTEXT

INCLUDE

DTID 3520 “Hemodynamic Clinical Context”

1

U

15

>>

CONTAINS

INCLUDE

DTID 3206 “VA Main Results”

1

M

16

>>

CONTAINS

INCLUDE

DTID 3207 “AA Main Results”

1

U

17

>>

CONTAINS

INCLUDE

DTID 3208 “Frame-to-Frame Results”

1

U

18

>>

CONTAINS

INCLUDE

DTID 3209 “Centerline Wall Motion”

1-2

U

VM = 1: Single plane analysis, VM = 2: Biplane analysis

19

>>

CONTAINS

INCLUDE

DTID 3210 “Radial Based Wall Motion”

1-2

U

VM = 1: Single plane analysis, VM = 2: Biplane analysis

20

>>

CONTAINS

INCLUDE

DTID 3211 “Landmark Based Wall Motion”

1-2

U

VM = 1: Single plane analysis, VM = 2: Biplane analysis


Content Item Descriptions

Row 7

Identifies the Ventricular Analysis program

Row 8

Identifies the Ventricular Analysis program version

Row 9

Identifies the Ventricular Analysis program manufacturer

Row 10

Identifies the ES and ED images on which the analysis is based, for frame by frame analysis the analyzed image are specified in the frame by frame results (3208) Template