DICOM PS3.16 2026d - Content Mapping Resource

Miscellaneous Ultrasound Report Templates 

TID 12000 General Ultrasound Report

This is the Template for the root of the Content Tree for a general ultrasound procedure report.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 12000. General Ultrasound Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

BCID 12320 “General Ultrasound Report Document Title”

1

M

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

U

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

M

4

>

CONTAINS

INCLUDE

DTID 12001 “Ultrasound Patient Characteristics”

1

U

5

>

CONTAINS

CONTAINER

DT (55111-9, LN, "Current Procedure Descriptions")

1

U

6

>>

CONTAINS

CODE

DT (125203, DCM, "Acquisition Protocol")

1-n

M

BCID 12001 “Ultrasound Protocol Type”

7

>>

CONTAINS

CODE

DT (113743, DCM, "Patient Orientation")

1

U

DCID 19 “Patient Orientation”

8

>>>

HAS CONCEPT MOD

CODE

EV (113744, DCM, "Patient Orientation Modifier")

1

U

DCID 20 “Patient Orientation Modifier”

9

>

CONTAINS

INCLUDE

DTID 1025 “Indications for Procedure”

1

U

$Findings = BCID 6051 “Breast Procedure Reason”

$Findings = BCID 12246 “Cardiac Ultrasound Indication for Study”

$Findings = BCID 12325 “Liver Ultrasound Study Indication”

12

>

CONTAINS

CONTAINER

DT (59776-5, LN, "Findings")

1-n

U

13

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

MC

IF Row 14 is not present

14

>>

CONTAINS

TEXT

EV (121071, DCM, "Finding")

1-n

MC

IF Row 13 is not present

15

>

CONTAINS

INCLUDE

DTID 5401 “Ultrasound Shear Wave Elastography Section”

1-n

U

16

>

CONTAINS

INCLUDE

DTID 12002 “Ultrasound Attenuation Coefficient Section”

1-n

U


Content Item Descriptions

Row 9

Baseline CIDs are provided for indications for Breast, Cardiac, and Liver procedures, but this does not preclude indications for other relevant ultrasound procedures.

Row 12

This container permits capturing findings that do not appropriately fit in one of the sections that follow.

DICOM PS3.16 2026d - Content Mapping Resource