DICOM PS3.16 2026d - Content Mapping Resource

Hemodynamics Report Templates 

The Templates that comprise the Hemodynamic Report are interconnected as shown in Figure A-6.

Hemodynamic Report Template Hierarchy

Figure A-6. Hemodynamic Report Template Hierarchy


Figure A-6 shows only the use of Templates specific to the Hemodynamic Report; it does not show common structural Templates such as TID 300 “Measurement”.

Note

Figure A-6 shows the relationship of Templates; it does not show the structural hierarchy of Content Items in the IOD. See Figure L-1 “Hemodynamics Report Structure” in PS3.17 .

TID 3500 Hemodynamics Report

The Hemodynamic Report Template is the root structure for the representation of measurements acquired during a procedure in a cardiac catheterization lab.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 3500. Hemodynamics Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122120, DCM, "Hemodynamics Report")

1

M

2

>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

M

3

>

INCLUDE

DTID 3601 “Procedure Context”

1

M

4

>

HAS OBS CONTEXT

INCLUDE

DTID 3602 “Cardiovascular Patient Characteristics”

1

M

5

>

HAS ACQ CONTEXT

INCLUDE

DTID 3603 “Procedure Environmental Characteristics”

1

U

6

>

CONTAINS

INCLUDE

DTID 3501 “Hemodynamics Measurement Group”

1-n

M

7

>

CONTAINS

INCLUDE

DTID 3570 “Summary, Hemodynamics”

1

U


DICOM PS3.16 2026d - Content Mapping Resource