Enhanced X-Ray Radiation Dose SR IOD Templates 

The Templates that comprise the Enhanced X-Ray Radiation Dose SR are interconnected as in Figure A-18b.

Enhanced X-Ray Radiation Dose Structured Report IOD Template Structure

Figure A-18b. Enhanced X-Ray Radiation Dose Structured Report IOD Template Structure


TID 10040 Enhanced X-Ray Radiation Dose

This template defines a container (the root) with subsidiary Content Items for documenting the radiation output for an ionizing radiation imaging device. This template supports cone-beam CT, in addition to other modalities.

Type:

Extensible

Order:

Non-Significant

Root:

Yes

Table TID 10040. Enhanced X-Ray Radiation Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113701, DCM, "X-Ray Radiation Dose Report")

1

M

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

3

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure reported")

1

M

BCID 10005 “X-Ray Radiation Dose Procedure Type Reported”

4

>>

HAS CONCEPT MOD

CODE

EV (363703001, SCT, "Has Intent")

1

M

DCID 3629 “Procedure Intent”

5

>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

M

6

>

HAS OBS CONTEXT

CODE

EV (113705, DCM, "Scope of Accumulation")

1

M

DCID 10000 “Scope of Accumulation”

7

>

CONTAINS

INCLUDE

DTID 10041 “Accumulated Dose Data”

1-n

U

8

>

CONTAINS

INCLUDE

DTID 10042 “Irradiation Event Summary Data”

1-n

M

9

>

CONTAINS

INCLUDE

DTID 10043 “Irradiation Details”

1

M

10

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U

11

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1

U

$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing")

12

>

CONTAINS

CODE

EV (113854, DCM, "Source of Dose Information")

1-n

M

DCID 10020 “Source of Projection X-Ray Dose Information”


Content Item Descriptions

Row 3

This row describes the general type of radiation-producing procedure from which the content of the RDSR is generated. More specific procedure codes, such as those used for ordering or scheduling, which may affect subsequent processing, may be recorded in the top level Data Set in Procedure Code Sequence (0008,1032) of the General Study Module.

Row 7

TID 10041 may be included once for each source. Summary-level values that may include multiple sources, e.g., Dose Area Product Total from planes A and B of a biplane system,could be described by an instance of TID 10041 where the Identification of X-Ray Source indicates multiple sources, e.g., "A and B".

Row 8

TID 10042 shall be included once for each irradiation event. Irradiation event values that may include multiple sources, e.g., CTDI for a dual-source CT scanner,could be described by an instance of TID 10042 where the Identification of X-Ray Source indicates multiple sources, e.g., "1 and 2".

DICOM PS3.16 2026d - Content Mapping Resource