DICOM PS3.16 2024d - Content Mapping Resource

C Acquisition Context Module, Protocol and Workflow Context Templates (Normative)

This Annex specifies the content of Templates for Acquisition, Protocol and Workflow Context required by DICOM IODs.

Templates for Acquisition, Protocol and Workflow Context 

TID 3401 ECG Acquisition Context

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 3401. ECG Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (10:11345, MDC, "Lead System")

1

U

BCID 3263 “Electrode Placement Value”

2

CODE

DT (109054, DCM, "Patient State")

1

U

BCID 3262 “ECG Patient State Value”

3

NUMERIC

DT (109055, DCM, "Protocol Stage")

1

U

UNITS = EV ({stage}, UCUM, "stage")

4

CODE

DT (109056, DCM, "Stress Protocol")

1

U

BCID 3261 “Stress Protocol”

5

NUMERIC

DCID 3690 “ECG Control Numeric Variable”

1-n

U

6

TEXT

DCID 3691 “ECG Control Text Variable”

1-n

U


TID 3403 Catheterization Acquisition Context

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 3403. Catheterization Acquisition Context


Note

See TID 3100 “Procedure Action” in Annex A for description of Procedure Action ID used in Row 5.

TID 3450 Cardiac Electrophysiology Acquisition Context

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 3450. Cardiac Electrophysiology Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (109061, DCM, "EP Procedure Phase")

1

U

BCID 3254 “Electrophysiology Procedure Phase”

2

NUM

EV (109060, DCM, "Procedure Step Number")

1

U

UNITS = EV ({step}, UCUM, "step")

3

TEXT

EV (109063, DCM, "Pulse train definition")

1

U


TID 3460 Projection Radiography Acquisition Context

Type:

Extensible

Order:

Non-Significant

Root:

No


TID 3470 NM/PET Acquisition Context

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 3470. NM/PET Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (109054, DCM, "Patient State")

1

M

DCID 3101 “Cardiac Procedural State Value”

2

INCLUDE

BTID 3471 “PET Covariates Acquisition Context”

1

U


TID 3471 PET Covariates Acquisition Context

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 3471. PET Covariates Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUMERIC

(14749-6, LN, "Glucose")

1

U

UNITS = EV (mmol/l, UCUM, "mmol/l")

2

DATE

(127857, DCM, "Glucose Measurement Date")

1

MC

IFF Row 1 is present and does not contain Observation DateTime (0040,A032)

3

TIME

(127858, DCM, "Glucose Measurement Time")

1

MC

IFF Row 1 is present and does not contain Observation DateTime (0040,A032)


Content Item Descriptions

Row 2

Glucose Measurement Date

In an earlier edition of the Standard, an incorrect DCM code was used for this concept, which was already assigned as (109081, DCM, "Prospective gating").

Row 3

Glucose Measurement Time

In an earlier edition of the Standard, an incorrect DCM code was used for this concept, which was already assigned as (109082, DCM, "Retrospective gating").

TID 3480 Neurophysiologic Stimulation Acquisition Context

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 3480. Neurophysiologic Stimulation Acquisition Context

NL

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (130491, DCM, "Stimulation Mode")

1

M

BCID 3041 “Neurophysiologic Stimulation Mode”

2

>

NUMERIC

EV (130492, DCM, "Stimulus Sample Position")

1

U

UNITS = EV (1, UCUM, "no units")

3

>

NUMERIC

EV (130493, DCM, "Stimulus Time Offset")

1

U

UNITS = DT (ms, UCUM, "ms")

4

>

NUMERIC

EV (130494, DCM, "Number of Stimulus Events")

1

U

UNITS = EV (1, UCUM, "no units")

5

>

NUMERIC

EV (130495, DCM, "Frequency of Stimulus Events")

1

MC

IFF Row 4 is present and contains a number greater than 1

UNITS = DT (Hz, UCUM, "Hz")


Content Item Descriptions

Row 2

Stimulus Sample Position

Sample number, where the first stimulation event takes place.

Note

The Stimulus Sample Position is not the number of the sample in the Waveform Data, but refers to the sample position in the multiplexed channels. E.g., with 4 multiplexed channels, the 10th sample in the Waveform Data is the 3rd sample in the 2nd channel, but Stimulus Sample Position represents the time of the 10th sample of the multiplexed data, i.e., corresponding to samples 37-40 in the Waveform Data.

Row 3

Stimulus Time Offset

The time offset, where the stimulation starts relative to the start time of the recorded waveform data.

Row 4

Number of Stimulus Events

If stimulation occurs periodically, this concept defines the number of applied stimuli.

Row 5

Frequency of Stimulus Events

If stimulation occurs periodically, this concept defines the frequency with which stimuli are applied.

TID 8001 Specimen Preparation

This Template describes a single specimen preparation step.

Type:

Extensible

Order:

Significant

Root:

No


Content Item Descriptions

Row 1

For sampling steps (which create a child specimen from a parent), the ID is that of the child specimen. For other preparation steps, the ID of a specimen does not change during the processing.

Row 2

The issuer shall be formatted in accordance with the HL7v2 Hierarchic Designator Data Type. That format is [ Namespace ID]^[ Universal ID^ Universal ID Type], where Namespace IDidentifies an entity within the local namespace or domain, Universal IDis a universal or unique identifier for an entity, and Universal ID Typespecifies the standard format of the Universal ID (see HL7 v2 Section 2.A.33).

Row 2a

When the specimen container is a slide, this information is present in the Container Type Code Sequence (0040,0518) in the Section C.7.6.22 “Specimen Module” in PS3.3 , and does not need to be repeated here.

Row 2b

When the specimen type is a slide, this information is present in the Specimen Type Code Sequence (0040,059A) in the Section C.7.6.22 “Specimen Module” in PS3.3 , and does not need to be repeated here.

TID 8002 Specimen Sampling

Type:

Extensible

Order:

Significant

Root:

No


Content Item Descriptions

Row 3

The Issuer of Specimen Identifier shall be formatted in accordance with the HL7 v2 Hierarchic Designator data type (see HL7 v2.6 Section 2.A.33), i.e., [ Namespace ID]^[ Universal ID^Universal ID Type]

Row 5

Description of coordinate system and origin reference point on parent specimen or parent specimen container used for localizing the sampling site

Rows 7, 8, 9

The X, Y and Z locations are used as needed to describe the sampling site; not all may be needed. E.g., resection from 10 cm along the colon may be described as only a Y dimension location.

Row 10

Reference to image of parent specimen localizing the sampling site; may include referenced Presentation State object

TID 8003 Specimen Staining

Type:

Extensible

Order:

Significant

Root:

No

Table TID 8003. Specimen Staining

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (424361007, SCT, "Using substance")

1-n

MC

IF Row 2 not present

DCID 8112 “Specimen Stain”

2

TEXT

DT (424361007, SCT, "Using substance")

1

MC

IF Row 1 not present


TID 8004 Specimen Localization

Type:

Extensible

Order:

Significant

Root:

No

Table TID 8004. Specimen Localization

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

TEXT

DT (111708, DCM, "Position Frame of Reference")

1

U

2

TEXT

DT (111718, DCM, "Location of Specimen")

1

U

3

NUMERIC

DT (111719, DCM, "Location of Specimen X offset")

1

U

4

NUMERIC

DT (111720, DCM, "Location of Specimen Y offset")

1

U

5

NUMERIC

DT (111721, DCM, "Location of Specimen Z offset")

1

U

6

IMAGE

DT (111718, DCM, "Location of Specimen")

1

U

7

COMPOSITE

DT (111718, DCM, "Location of Specimen")

1

U

Presentation State SOP Instance reference

8

TEXT

DT (111723, DCM, "Visual Marking of Specimen")

1

U


Content Item Descriptions

Row 1

Description of coordinate system and origin reference point used for localizing the Specimen. The value "CURRENT IMAGE " identifies the frame of reference as the pixel space of the Image SOP Instance in which this Content Item occurs.

Row 2

Description of specimen location, either in absolute terms or relative to the Position Frame Reference of Row 1

Rows 3, 4, 5

Location of specimen (nominal center) relative to the Position Frame Reference of Row 1. The Content Items include the units of measurement (e.g., mm). If Row 1 value is "CURRENT IMAGE ", measurement shall be from the top left hand corner of the Pixel Data of the SOP Instance, using units of ({pixel}, UCUM, "Pixels").

Row 6

Reference to image of container localizing the specimen; may include referenced Presentation State object

Row 7

Reference to Presentation State object for this SOP Instance, with annotations localizing the specimen

Row 8

Description of visual distinguishing identifiers, e.g., ink, or a particular shape of the specimen

TID 8010 Slide Imaging Parameters

This Template describes protocol parameters for a Slide Imaging Procedure Step. As an extensible Template, additional items may be included using other concept names from standard or private coding schemes.

Type:

Extensible

Order:

Significant

Root:

No


TID 8200 Radiology Reading Task Parameters

This Template describes parameters for a radiology reading task.

Note

Specialty to Read is nested inside Modality to Read in order to facilitate C-FIND matching against both Modality and Specialty.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 8200. Radiology Reading Task Parameters


TID 8300 Skin Imaging Acquisition Context

This Template provides defines an Acquisition Context Template for skin imaging. The attributes in this template represent values known at the time of image acquisition. Hence, these values may subsequently change.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 8300. Skin Imaging Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (443635002, SCT, "Fitzpatrick Skin Type")

1

U

DCID 4401 “Fitzpatrick Skin Type”

2

CODE

EV (415229000, SCT, "Racial group")

1

U

BCID 6099 “Racial Group”

3

CODE

EV (161432005, SCT, "History of malignant melanoma")

1-n

U

DCID 4402 “History of Malignant Melanoma”

4

NUMERIC

DT (130483, DCM, "Number of malignant melanomas")

1

UC

IFF Row 3 is present

5

CODE

EV (1251000119106, SCT, "History of melanoma in situ of skin")

1-n

U

DCID 4403 “History of Melanoma in Situ”

6

NUMERIC

DT (130484, DCM, "Number of melanomas in situ")

1

UC

IFF Row 5 is present

7

CODE

EV (130482, DCM, "History of non-melanoma skin cancer")

1-n

U

DCID 4404 “History of Non-Melanoma Skin Cancer”

8

CODE

EV (64572001, SCT, "Disease")

1-n

U

DCID 4405 “Skin Disorder”

9

CODE

EV (427858005, SCT, "Family history of malignant melanoma")

1-n

U

DCID 4402 “History of Malignant Melanoma”

10

NUMERIC

DT (130487, DCM, "Number of first-degree relatives affected by malignant melanoma")

1

UC

IFF Row 9 is present

11

CODE

EV (130481, DCM, "Family history of melanoma in situ")

1-n

U

DCID 4403 “History of Melanoma in Situ”

12

CODE

EV (130480, DCM, "Family history of non-melanoma skin cancer")

1-n

U

DCID 4403 “History of Melanoma in Situ”

13

CODE

EV (418799008, SCT, "Findings reported by patient/informant")

1-n

U

DCID 4406 “Patient Reported Lesion Characteristic”

14

CODE

EV (118242002, SCT, "Finding by palpation")

1-n

U

DCID 4407 “Lesion Palpation Finding”

15

CODE

EV (118243007, SCT, "Finding by inspection")

1-n

U

DCID 4408 “Lesion Visual Finding”

16

CODE

EV (416940007, SCT, "Past history of procedure")

1-n

U

DCID 4409 “Skin Procedure”

17

CODE

EV (130832, DCM, "Skin lesion color")

1-n

U

DCID 4411 “Lesion Color”

18

CODE

EV (386439008, SCT, "Skin care topical treatments")

1-n

U

DCID 4410 “Topical Treatment”

19

CODE

EV (C4684549, NCIt, "New Lesion Indicator")

1

U

DCID 230 “Yes-No”


Content Item Descriptions

Row 13

Finding reported by patient/informant prior to imaging.

Row 18

Recent topical treatments relevant to this imaging acquisition.

TID 8301 Specimen Staining for Cutaneous Confocal Microscopy

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 8301. Specimen Staining for Cutaneous Confocal Microscopy

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (424361007, SCT, "Using substance")

1-n

MC

IF Row 2 not present

DCID 4412 “Specimen Stain for Confocal Microscopy”

2

TEXT

DT (424361007, SCT, "Using substance")

1

MC

IF Row 1 not present


TID 15100 Contrast Agent/Pre-Medication Protocol Context

This Template specifies medications to be administered prior to a diagnostic imaging protocol, imaging contrast agents to be used in the protocol, and/or bolus agents to be used in the protocol. Each medication or agent may be modified by a specified route of administration. The top level Content Items of this Template may appear in any order in the Protocol Context Sequence, hence the order in this Template is not significant. There may be significance in the order in which the Content Items are included in the Protocol Context Sequence, e.g., the requested order in which pre-medications are to be administered.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15100. Contrast Agent/Pre-Medication Protocol Context


TID 15101 NM/PET Protocol Context

Type:

Extensible

Order:

Significant

Root:

No

Table TID 15101. NM/PET Protocol Context

NL

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (349358000, SCT, "Radiopharmaceutical agent")

1

M

BCID 25 “Radiopharmaceutical”

BCID 4021 “PET Radiopharmaceutical”

2

>

CODE

EV (89457008, SCT, "Radionuclide)

1

U

BCID 18 “Radiopharmaceutical Isotope”

BCID 4020 “PET Radionuclide”

3

>

UIDREF

EV (113503, DCM, "Radiopharmaceutical Administration Event UID")

1

U

4

>

DATETIME

EV (123003, DCM, "Radiopharmaceutical Start DateTime")

1

U

5

>

DATETIME

EV (123004, DCM, "Radiopharmaceutical Stop DateTime")

1

U

6

>

NUMERIC

EV (123005, DCM, "Radiopharmaceutical Volume")

1

U

UNITS = DT (cm3, UCUM, "cm3")

7

>

NUMERIC

EV (123006, DCM, "Radionuclide Total Dose")

1

U

UNITS = DT (Bq, UCUM, "Bq")

8

>

NUMERIC

EV (123007, DCM, "Radiopharmaceutical Specific Activity")

1

U

UNITS = DT (Bq/mol, UCUM, "Bq/mol")

9

>

CODE

EV (410675002, SCT, "Route of Administration")

1

U

BCID 11 “Administration Route”

10

>

NUMERIC

EV (123009, DCM, "Radionuclide Syringe Counts")

1

U

UNITS = DT ({counts}/s, UCUM "counts/s")

11

>

NUMERIC

EV (123010, DCM, "Radionuclide Residual Syringe Counts")

1

U

UNITS = DT ({counts}/s, UCUM "counts/s")

12

NUMERIC

EV (14749-6, LN, "Glucose")

1

U

UNITS = EV (mmol/l, UCUM, "mmol/l")

13

>

DATE

EV (127857, DCM, "Glucose Measurement Date")

1

MC

IFF Row 12 is present and does not contain Observation DateTime (0040,A032)

14

>

TIME

EV (127858, DCM, "Glucose Measurement Time")

1

MC

IFF Row 12 is present and does not contain Observation DateTime (0040,A032)


Content Item Descriptions

Row 13

Glucose Measurement Date

In an earlier edition of the Standard, an incorrect DCM code was used for this concept, which was already assigned as (109081, DCM, "Prospective gating").

Row 14

Glucose Measurement Time

In an earlier edition of the Standard, an incorrect DCM code was used for this concept, which was already assigned as (109082, DCM, "Retrospective gating").

TID 15200 JJ1017 Protocol Context

This Template defines protocol context concepts to support the requirements of Japanese Guideline JJ1017. This is expected to be used with Scheduled or Performed Protocol Codes from Coding Scheme JJ1017-16M defined in Guideline JJ1017.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 15200. JJ1017 Protocol Context

NL

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (123016, DCM, "Imaging Conditions")

1

M

Baseline terms from Coding Scheme JJ1017-16S of JJ1017 version 3.0


TID 15300 RT Prescription Annotation

The concepts in this TID are topics of advice or information provided by the prescribing physician for planning, preparation and delivery of treatment for a prescription.

Type:

Extensible

Order:

Non-Significant

Root:

No


Content Item Descriptions

Rows 11, 12, 13, 14

The source of radiation to be used for this RT treatment. More than one source indicates that the RT treatment may use any combination for treatment. There is no defined relationship between the entries in Rows 11, 12, 13, 14 and entries in the Rows 15 and 16.

Row 15

Including several energies indicates that they may be used in any combination.

Row 16

The codes identify procedures supporting the patient positioning process prior to RT treatment. Including several procedures indicates that they may be used in any combination.

TID 15301 RT Segment Characteristics

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15301. RT Segment Characteristics

NL

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUMERIC

EV (130082, DCM, "Relative Mass Density")

1

U

UNITS = EV ({ratio}, UCUM, "ratio")

2

NUMERIC

EV (130083, DCM, "Relative Electron Density")

1

U

UNITS = EV ({ratio}, UCUM, "ratio")

3

NUMERIC

EV (130084, DCM, "Effective Z")

1

U

UNITS = EV (1, UCUM, "no units")

4

NUMERIC

EV (130085, DCM, "Effective Z per A")

1

U

UNITS = EV (/u, UCUM, "/u")

5

NUMERIC

EV (130086, DCM, "Relative Linear Stopping Power")

1

U

UNITS = EV ({ratio}, UCUM, "ratio")

6

>

NUMERIC

EV (130087, DCM, "Reference Energy")

1

M

UNITS = EV (MeV, UCUM, "Megaelectronvolt")

7

NUMERIC

EV (130088, DCM, "Linear Cell Kill Factor")

1

U

UNITS = EV ({ratio}, UCUM, "ratio")

8

NUMERIC

EV (130089, DCM, "Quadratic Cell Kill Factor")

1

U

UNITS = EV ({ratio}, UCUM, "ratio")

9

NUMERIC

EV (130090, DCM, "High Dose Fraction Linear Cell Kill Factor")

1

U

UNITS = EV ({ratio}, UCUM, "ratio")

10

NUMERIC

EV (130091, DCM, "Half-time for Tissue Repair")

1

U

UNITS = EV (s, UCUM, "second")

11

NUMERIC

EV (130092, DCM, "High Dose Fraction Transition Dose")

1

U

UNITS = EV (Gy, UCUM, "Gray")

12

NUMERIC

EV (130093, DCM, "Atomic Number")

1-n

U

UNITS = EV (1, UCUM, "no units")

13

>

NUMERIC

EV (130094, DCM, "Elemental Composition Atomic Mass Fraction")

1

M

UNITS = EV ({ratio}, UCUM, "ratio")

14

NUMERIC

EV (130095, DCM, "alpha gEUD value")

1

U

UNITS = EV ({ratio}, UCUM, "ratio")

15

CODE

EV (130737, DCM, "RT Segment Material")

1

U

BCID 9579 “RT Segment Material”


Content Item Descriptions

Rows 12, 13

The value of (130094, DCM, "Elemental Composition Atomic Mass Fraction") annotates the fractional weight of the elements identified by the (130093, DCM, "Atomic Number") with respect to the total mass of the segment. The allowed value is in the range of [0, 1].

TID 15302 Patient Support Position Parameters

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15302. Patient Support Position Parameters

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUMERIC

EV (126802, DCM, "IEC61217 Table Top Continuous Pitch Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

2

NUMERIC

EV (126803, DCM, "IEC61217 Table Top Continuous Roll Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

3

NUMERIC

EV (126801, DCM, "IEC61217 Patient Support Continuous Yaw Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

4

NUMERIC

EV (126804, DCM, "IEC61217 Table Top Eccentric Axis Distance")

1

U

UNITS = EV (mm, UCUM, "mm")

5

NUMERIC

EV (126805, DCM, "IEC61217 Table Top Continuous Eccentric Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

6

NUMERIC

EV (126806, DCM, "IEC61217 Table Top Lateral Position")

1

U

UNITS = EV (mm, UCUM, "mm")

7

NUMERIC

EV (126807, DCM, "IEC61217 Table Top Longitudinal Position")

1

U

UNITS = EV (mm, UCUM, "mm")

8

NUMERIC

EV (126808, DCM, "IEC61217 Table Top Vertical Position")

1

U

UNITS = EV (mm, UCUM, "mm")

9

NUMERIC

EV (126812, DCM, "Isocentric Patient Support Continuous Pitch Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

10

NUMERIC

EV (126813, DCM, "Isocentric Patient Support Continuous Roll Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

11

NUMERIC

EV (126814, DCM, "Isocentric Patient Support Continuous Yaw Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

12

NUMERIC

EV (126815, DCM, "Isocentric Patient Support Lateral Position")

1

U

UNITS = EV (mm, UCUM, "mm")

13

NUMERIC

EV (126816, DCM, "Isocentric Patient Support Longitudinal Position")

1

U

UNITS = EV (mm, UCUM, "mm")

14

NUMERIC

EV (126817, DCM, "Isocentric Patient Support Vertical Position")

1

U

UNITS = EV (mm, UCUM, "mm")


TID 15303 Radiotherapy Treatment Scheduled Processing Parameters

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15303. Radiotherapy Treatment Scheduled Processing Parameters

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

TEXT

EV (121384, DCM, "RT Plan Label")

1

U

2

NUMERIC

EV (121385, DCM, "Current Fraction Number")

1

U

UNIT = EV (1, UCUM, "no units")

3

NUMERIC

EV (121386, DCM, "Number of Fractions Planned")

1

U

UNIT = EV (1, UCUM, "no units")

4

NUMERIC

EV (121387, DCM, "Number of Fractions Completed")

1

U

UNIT = EV (1, UCUM, "no units")

5

CODE

EV (121388, DCM, "Checked-In Status")

1

U

DCID 230 “Yes-No”


TID 15304 Radiotherapy Treatment Progress Parameters

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15304. Radiotherapy Treatment Progress Parameters

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUMERIC

EV (121389, DCM, "Referenced Beam Number")

1

U

UNIT = EV (1, UCUM, "no units")


TID 15305 Patient Setup Fixation Device Parameters

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15305. Patient Setup Fixation Device Parameters

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

TEXT

EV (130657, DCM, "Couch Index Label")

1

U

2

NUMERIC

EV (130658, DCM, "Fixation Device Angle")

1

U

UNIT = EV (deg, UCUM, "deg")

3

NUMERIC

EV (130659, DCM, "Abdominal Compression Plate Position Number")

1

U

UNIT = EV (1, UCUM, "no units")

4

NUMERIC

EV (130660, DCM, "Abdominal Compression Belt Length")

1

U

UNIT = EV (mm, UCUM, "no mm")

5

NUMERIC

EV (130661, DCM, "Abdominal Compression Belt Pressure")

1

U

UNIT = EV (Pa, UCUM, "Pa")

6

NUMERIC

EV (130840, DCM, "Seat Pan Height")

1

U

UNIT = EV (mm, UCUM, "mm")

7

NUMERIC

EV (130841, DCM, "Seat Pan Pitch Angle")

1

U

UNIT = EV (deg, UCUM, "deg")

8

NUMERIC

EV (130842, DCM, "Backrest Fixation Pitch Angle")

1

U

UNIT = EV (deg, UCUM, "deg")

9

NUMERIC

EV (130843, DCM, "Shin Rest Fixation Position")

1

U

UNIT = EV (mm, UCUM, "mm")

10

NUMERIC

EV (130844, DCM, "Heel Fixation Stop Position")

1

U

UNIT = EV (mm, UCUM, "mm")

11

NUMERIC

EV (130845, DCM, "Left Arm Rest Position")

1

U

UNIT = EV (mm, UCUM, "mm")

12

NUMERIC

EV (130846, DCM, "Left Arm Rest Pitch Angle")

1

U

UNIT = EV (deg, UCUM, "deg")

13

NUMERIC

EV (130847, DCM, "Left Arm Rest Roll Angle")

1

U

UNIT = EV (deg, UCUM, "deg")

14

NUMERIC

EV (130848, DCM, "Right Arm Rest Position")

1

U

UNIT = EV (mm, UCUM, "mm")

15

NUMERIC

EV (130849, DCM, "Right Arm Rest Pitch Angle")

1

U

UNIT = EV (deg, UCUM, "deg")

16

NUMERIC

EV (130850, DCM, "Right Arm Rest Roll Angle")

1

U

UNIT = EV (deg, UCUM, "deg")

17

CODE

EV (130851, DCM, "Hand Grips Presence")

1

U

DCID 240 “Present-Absent”


Content Item Descriptions

Row 2

The values of these rows refer to angles, which are device-specific and are not defined based on a standardized coordinate system. are device-specific and are not defined based on a standardized coordinate system.

TID 15306 Patient Setup Alignment Device Parameters

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15306. Patient Setup Fixation Alignment Parameters

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

TEXT

EV (130657, DCM, "Couch Index Label")

1

U

2

COMPOSITE

EV (130662, DCM, "Referenced Patient Alignment Reference")

1-n

U

3

CODE

EV (130666, DCM, "Radiotherapy Fiducial")

1-n

U

BCID 7112 “Radiotherapy Fiducial”


Content Item Descriptions

Row 2

Composite Instances to be used by a treatment setup device, such as a reference surface mesh to be used by an optical surface scanner.

Row 3

Fiducials used in radiotherapy procedures e.g., as a reference location during laser setup.

TID 15307 Acquisition Initiation Parameters

This Template specifies how an acquisition of data by devices, which perform such operations to construct images or similar objects (like surfaces, fiducials, etc.), is to be initiated. Acquisitions may be initiated manually or triggered when parameters have specified values. Parameterized acquisitions may be initiated once or repeatedly.

This Template does not contain patient-related parameters; however, the Template is extensible and implementations may include such parameters (e.g., a measurement of a certain value of distance between a device and the patient surface triggering the acquisition).

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15307. Acquisition Initiation Parameters


Content Item Descriptions

Rows 3, 4, 5, 6

These rows provide the value(s) of a triggering parameter associated with acquisition initiation.

If Row 2 has the value of (373066001, SCT, "Yes") these rows contain 2 or 3 values. Value 1 represents the start value at which the repeated acquisition starts and value 2 contains the interval value by which the start value is repeatedly incremented to trigger subsequent acquisitions. Value 3 if present contains the stop value at which, when exceeded, no further acquisitions are started. If the stop value is not provided, the acquisition is triggered until the end of the therapeutic radiation.

If Row 2 has the value of (373067005, SCT, "No") these rows shall contain a distinct value for the start of each acquisition. If more than one value is present, the values shall be specified in increasing order.

TID 15308 Imaging Source Geometry Parameters

This Template specifies the location and orientation of an imaging source coordinate system.

For an IEC 61217 GANTRY coordinate system, the Z-axis passes through, and is directed towards, the radiation source, which in this context is the imaging source.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15308. Imaging Source Geometry Parameters

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUMERIC

EV (126809, DCM, "IEC61217 Gantry Continuous Roll Angle")

1

U

Units = EV (deg, UCUM, "deg")

2

NUMERIC

EV (126810, DCM, "IEC61217 Gantry Continuous Pitch Angle")

1

U

Units = EV (deg, UCUM, "deg")

3

NUMERIC

EV (126811, DCM, "IEC61217 Gantry Continuous Yaw Angle")

1

U

Units = EV (deg, UCUM, "deg")

4

NUMERIC

EV (130801, DCM, "IEC61217 Imaging Source to Axis Distance")

1

U

Units = EV (mm, UCUM, "mm")


TID 15309 Image Receptor Geometry Parameters

This Template specifies the location and orientation of an image receptor coordinate system.

For an IEC 61217 GANTRY coordinate system, the Z-axis passes through, and is directed towards, the radiation source, which in this context is the imaging source.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15309. Image Receptor Geometry Parameters

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUMERIC

EV (126809, DCM, "IEC61217 Gantry Continuous Roll Angle")

1

U

Units = EV (deg, UCUM, "deg")

2

NUMERIC

EV (126810, DCM, "IEC61217 Gantry Continuous Pitch Angle")

1

U

Units = EV (deg, UCUM, "deg")

3

NUMERIC

EV (126811, DCM, "IEC61217 Gantry Continuous Yaw Angle")

1

U

Units = EV (deg, UCUM, "deg")

4

NUMERIC

EV (130802, DCM, "IEC61217 X-Ray Image Receptor Radial Displacement from Isocenter")

1

U

Units = EV (mm, UCUM, "mm")

5

NUMERIC

EV (130803, DCM, "IEC61217 X-Ray Image Receptor Longitudinal Displacement")

1

U

Units = EV (mm, UCUM, "mm")

6

NUMERIC

EV (130804, DCM, "IEC61217 X-Ray Image Receptor Lateral Displacement")

1

U

Units = EV (mm, UCUM, "mm")

7

NUMERIC

EV (130805, DCM, "IEC61217 X-Ray Image Receptor Rotation")

1

U

Units = EV (deg, UCUM, "deg")


Change the following code definitions in PS3.16, Annex D:

TID 15310 Imaging Device Position Parameters

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15310. Imaging Device Position Parameters

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUMERIC

EV (126810, DCM, "IEC61217 Gantry Continuous Pitch Angle")

1

U

Units = EV (deg, UCUM, "deg")

1

NUMERIC

EV (126811, DCM, "IEC61217 Gantry Continuous Yaw Angle")

1

U

Units = EV (deg, UCUM, "deg")


TID 15400 Real-World Quantity Definition

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15400. Real-World Quantity Definition


Content Item Descriptions

Row 1

This row uses a concept name that specifies the quantified characteristic. It is not required that (246205007, SCT, "Quantity") be used if there is a reason to use a similar concept.

Row 2

May be concept modifiers, such as (370129005, SCT, "Measurement Method").

TID 15401 Real-World Quantity Definition for X-Ray Attenuation Properties

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 15401. Real-World Quantity Definition for X-Ray Attenuation Properties

NL

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

INCLUDE

DTID 15400 “Real-World Quantity Definition”

1

M

2

NUMERIC

EV (130087, DCM, "Reference Energy")

1

MC

IF TID 15400 Row 1 Quantity value is (130086, DCM, "Relative Linear Stopping Power")

UNITS = EV ("MeV", UCUM, "Megaelectronvolt")


DICOM PS3.16 2024d - Content Mapping Resource