DICOM PS3.15 2021e - Security and System Management Profiles

A.5.3.7 DICOM Instances Transferred

This message describes the event of the completion of transferring DICOM SOP Instances between two Application Entities. This message may only include information about a single patient.

Note

This message may have been preceded by a Begin Transferring Instances message. The Begin Transferring Instances message conveys the intent to store SOP Instances, while the Instances Transferred message records the completion of the transfer. Any disagreement between the two messages might indicate a potential security breach.

Table A.5.3.7-1. Audit Message for DICOM Instances Transferred

Real World Entities

Field Name

Opt.

Value Constraints

Event

EventID

M

EV (110104, DCM, "DICOM Instances Transferred")

EventActionCode

M

Enumerated Value:

C = (create) if the receiver did not hold copies of the instances transferred

R = (read) if the receiver already holds copies of the SOP Instances transferred, and has determined that no changes are needed to the copies held.

U = (update) if the receiver is altering its held copies to reconcile differences between the held copies and the received copies.

If the Audit Source is either not the receiver, or otherwise does not know whether or not the instances previously were held by the receiving node, then use "R" = (Read).

EventDateTime

M

Shall be the time when the transfer has completed

EventOutcomeIndicator

M

not specialized

EventTypeCode

U

not specialized

Active Participant:

Process that sent the data (1)

UserID

M

not specialized

AlternativeUserID

U

not specialized

UserName

U

not specialized

UserIsRequestor

M

not specialized

RoleIDCode

M

EV (110153, DCM, "Source Role ID")

NetworkAccessPointTypeCode

U

not specialized

NetworkAccessPointID

U

not specialized

Active Participant:

The process that received the data. (1)

UserID

M

not specialized

AlternativeUserID

U

not specialized

UserName

U

not specialized

UserIsRequestor

M

not specialized

RoleIDCode

M

EV (110152, DCM, "Destination Role ID")

NetworkAccessPointTypeCode

U

not specialized

NetworkAccessPointID

U

not specialized

Active Participant:

Other participants that are known, especially third parties that are the requestor (0..N)

UserID

M

not specialized

AlternativeUserID

U

not specialized

UserName

U

not specialized

UserIsRequestor

M

not specialized

RoleIDCode

U

not specialized

NetworkAccessPointTypeCode

U

not specialized

NetworkAccessPointID

U

not specialized

Participating Object:

Studies being transferred (1..N)

ParticipantObjectTypeCode

M

Shall be: 2 = system

ParticipantObjectTypeCodeRole

M

Shall be: 3 = report

ParticipantObjectDataLifeCycle

U

not specialized

ParticipantObjectIDTypeCode

M

EV (110180, DCM, "Study Instance UID")

ParticipantObjectSensitivity

U

not specialized

ParticipantObjectID

M

The Study Instance UID

ParticipantObjectName

U

not specialized

ParticipantObjectQuery

U

not specialized

ParticipantObjectDetail

U

Not specialized

ParticipantObjectDescription

U

Not specialized

SOPClass

MC

See Table A.5.2-1

Accession

U

not specialized

NumberOfInstances

U

not specialized

Instances

U

not specialized

Encrypted

U

not specialized

Anonymized

U

not specialized

Participating Object:

Patient (1)

ParticipantObjectTypeCode

M

Shall be: 1 = person

ParticipantObjectTypeCodeRole

M

Shall be: 1 = patient

ParticipantObjectDataLifeCycle

U

not specialized

ParticipantObjectIDTypeCode

M

Shall be: EV (2, RFC-3881, "Patient Number")

ParticipantObjectSensitivity

U

not specialized

ParticipantObjectID

M

The patient ID

ParticipantObjectName

U

The patient name

ParticipantObjectQuery

U

not specialized

ParticipantObjectDetail

U

not specialized

ParticipantObjectDescription

U

not specialized


DICOM PS3.15 2021e - Security and System Management Profiles