DICOM PS3.15 2022b - Security and System Management Profiles

A.5.3.12 User Authentication

This message describes the event that a user has attempted to log on or log off. This report can be made regardless of whether the attempt was successful or not. No Participant Objects are needed for this message.

Note

The user usually has UserIsRequestor TRUE, but in the case of a logout timer, the Node might be the UserIsRequestor.

Table A.5.3.12-1. Audit Message for User Authentication

Real World Entities

Field Name

Opt.

Value Constraints

Event

EventID

M

EV (110114, DCM, "User Authentication")

EventActionCode

M

Shall be: E = Execute

EventDateTime

M

not specialized

EventOutcomeIndicator

M

not specialized

EventTypeCode

M

Defined Terms:

EV (110122, DCM, "Login")

EV (110123, DCM, "Logout")

Active Participant:

Person Authenticated or claimed

(1)

UserID

M

not specialized

AlternativeUserID

U

not specialized

UserName

U

not specialized

UserIsRequestor

M

not specialized

RoleIDCode

U

not specialized

NetworkAccessPointTypeCode

M

not specialized

NetworkAccessPointID

M

not specialized

Active Participant:

Node or System performing authentication (0..1)

UserID

M

not specialized

AlternativeUserID

U

not specialized

UserName

U

not specialized

UserIsRequestor

M

not specialized

RoleIDCode

U

not specialized

NetworkAccessPointTypeCode

U

not specialized

NetworkAccessPointID

U

not specialized


DICOM PS3.15 2022b - Security and System Management Profiles