DICOM PS3.15 2026d - Security and System Management Profiles

A.5.3.8 DICOM Study Deleted

This message describes the event of deletion of one or more Studies and all associated SOP Instances in a single action. This message shall only include information about a single patient.

Table A.5.3.8-1. Audit Message for DICOM Study Deleted

Real-World Entities

Field Name

Opt.

Value Constraints

Event: EventIdentification

EventID

M

EV (110105, DCM, "DICOM Study Deleted")

EventActionCode

M

Shall be:

D

Delete

EventDateTime

M

Not specialized.

EventOutcomeIndicator

M

Not specialized.

EventTypeCode

U

Not specialized.

Active Participant: ActiveParticipant

The person and/or process deleting the Study (1..2)

UserID

M

Not specialized.

AlternativeUserID

U

Not specialized.

UserName

U

Not specialized.

UserIsRequestor

M

Not specialized.

RoleIDCode

U

Not specialized.

NetworkAccessPointTypeCode

U

Not specialized.

NetworkAccessPointID

U

Not specialized.

Participant Object: ParticipantObjectIdentification

Studies being deleted (1..N)

ParticipantObjectTypeCode

M

Shall be:

2

System Object

ParticipantObjectTypeCodeRole

M

Shall be:

3

Report

ParticipantObjectDataLifeCycle

U

Not specialized.

ParticipantObjectIDTypeCode

M

EV (110180, DCM, "Study Instance UID")

ParticipantObjectSensitivity

U

Not specialized.

ParticipantObjectID

M

The Study Instance UID.

ParticipantObjectName

MC

Required if ParticipantObjectQuery is not present.

ParticipantObjectQuery

MC

Required if ParticipantObjectName is not present.

ParticipantObjectDetail

U

Not specialized.

ParticipantObjectDescription

U

Not specialized.

SOPClass

MC

See Section A.5.2.

Accession

U

Not specialized.

NumberOfInstances

U

Not specialized.

Instances

U

Not specialized.

Encrypted

U

Not specialized.

Anonymized

U

Not specialized.

Participant Object: ParticipantObjectIdentification

Patient (1)

ParticipantObjectTypeCode

M

Shall be:

1

Person

ParticipantObjectTypeCodeRole

M

Shall be:

1

Patient

ParticipantObjectDataLifeCycle

U

Not specialized.

ParticipantObjectIDTypeCode

M

Shall be:

2

Patient Number

ParticipantObjectSensitivity

U

Not specialized.

ParticipantObjectID

M

The patient ID.

ParticipantObjectName

M

The patient name.

ParticipantObjectDetail

U

Not specialized.

ParticipantObjectDescription

U

Not specialized.


DICOM PS3.15 2026d - Security and System Management Profiles