DICOM PS3.11 2021e - Media Storage Application Profiles

N.3 STD-GEN-BD-MPEG4-LV42 and STD-GEN-SEC-BD-MPEG4-LV42 Profile Classes

N.3.1 SOP Classes and Transfer Syntaxes

This Application Profile is based on the Media Storage Service Class with the Interchange Option (see PS3.4).

Table N.3-1. STD-GEN-BD-MPEG4-LV42 and STD-GEN-SEC-BD-MPEG4-LV42 SOP Classes and Transfer Syntaxes

Information Object Definition

SOP Class UID

Transfer Syntax and UID

FSC Requirement

FSR Requirement

FSU Requirement

Basic Directory

1.2.840.10008.1.3.10

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Mandatory

Mandatory

Mandatory

Multi-frame Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

MPEG-4 AVC/H.264 High Profile / Level 4.2 For 2D Video

1.2.840.10008.1.2.4.104

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Multi-frame Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

MPEG-4 AVC/H.264 High Profile / Level 4.2 For 3D Video

1.2.840.10008.1.2.4.105

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Multi-frame Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

MPEG-4 AVC/H.264 Stereo High Profile / Level 4.2

1.2.840.10008.1.2.4.106

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement


The SOP Classes and corresponding Transfer Syntax supported by this Application Profile are specified in the Table N.3-1. The supported Storage SOP Class(es) shall be listed in the Conformance Statement using a table of the same form.

DICOM PS3.11 2021e - Media Storage Application Profiles