DICOM PS3.11 2021e - Media Storage Application Profiles

D.3.3 Directory Information in DICOMDIR

Conformant Application Entities shall include in the DICOMDIR File the Basic Directory IOD containing Directory Records at the Patient and the subsidiary Study and Series levels, appropriate to the SOP Classes in the File Set.

All DICOM files in the File Set incorporating SOP Instances defined for the specific Application Profile shall be referenced by Directory Records.

Note

DICOMDIRs with no directory information are not allowed by this Application Profile.

All implementations shall include the DICOM Media Storage Directory in the DICOMDIR file. There shall only be one DICOMDIR file per File Set. The DICOMDIR file shall be in the root directory of the medium. The Patient ID at the patient level shall be unique for each patient directory record in one File Set.

D.3.3.1 Additional Keys

File Set Creators and Updaters are required to generate the mandatory elements specified in PS3.3.

Table D.3-2 specifies the additional associated keys. At each directory record level other additional data elements can be added, but it is not required that File Set Readers be able to use them as keys. Refer to the Basic Directory IOD in PS3.3.

Table D.3-2. STD-GEN Additional DICOMDIR Keys

Key Attribute

Tag

Directory Record Type

Type

Notes

Image Type

(0008,0008)

IMAGE

1C

Required if present in image object.

Referenced Image Sequence

(0008,1140)

IMAGE

1C

Required if present in image object.

>Referenced SOP Class UID

(0008,1150)

IMAGE

1C

Required if Referenced Image Sequence (0008,1140) is present

>Referenced SOP Instance UID

(0008,1155)

IMAGE

1C

Required if Referenced Image Sequence (0008,1140) is present.

>All other elements from Referenced Image Sequence (including Purpose of Reference Code Sequence and its content)

IMAGE

3


Note

The requirements with respect to the mandatory DICOMDIR keys in PS3.3 imply that either these attributes are present in the Image IOD, or they are in some other way supplied by the File-set Creator. These attributes are (0010,0020) Patient ID, (0008,0020) Study Date, (0008,0030) Study Time, (0020,0010) Study ID, (0020,0011) Series Number, and (0020,0013) Instance Number.

DICOM PS3.11 2021e - Media Storage Application Profiles