| DICOM PS3.11 2026b - Media Storage Application Profiles |
|---|
Conformant Application Entities shall include in the DICOMDIR File the Basic Directory IOD containing Directory Records at the Patient and the subsidiary Study and Series levels, appropriate to the SOP Classes in the File-set.
All DICOM files in the File-set incorporating SOP Instances defined for the specific Media Storage Application Profile shall be referenced by Directory Records.
DICOMDIRs with no directory information are not allowed by this Media Storage Application Profile.
All implementations shall include the DICOM Media Storage Directory in the DICOMDIR file. There shall only be one DICOMDIR file per File-set. The DICOMDIR file shall be in the root directory of the medium. The Patient ID at the patient level shall be unique for each patient directory record in one File-set.
File-set Creators and File-set Updaters are required to generate the required Attributes specified in PS3.3.
Table H.3-2 specifies the additional associated keys that shall also be applicable to the profiles defined in this Annex. At each directory record level other additional Data Elements can be added, but it is not required that File-set Readers be able to use them as keys. Refer to the Basic Directory IOD in PS3.3.
| DICOM PS3.11 2026b - Media Storage Application Profiles |
|---|