DICOM PS3.11 2026c - Media Storage Application Profiles

L.3.3 Directory Information in DICOMDIR

The Directory shall include Directory Records of PATIENT, STUDY, SERIES, IMAGE corresponding to the information object files in the File-set. All DICOM files in the File-set incorporating SOP Instances (Information Objects) defined for the specific Media Storage Application Profile shall be referenced by Directory Records.

Note

DICOMDIRs with no directory information are not allowed by these Media Storage Application Profiles.

There may only be one DICOMDIR file per File-set. The Patient ID at the patient level shall be unique for each patient directory record in one File-set.

L.3.3.1 Additional Keys

No additional keys are specified.

DICOM PS3.11 2026c - Media Storage Application Profiles