DICOM PS3.11 2026a - Media Storage Application Profiles

H.3 STD-GEN-DVD and STD-GEN-SEC-DVD Profile Classes

H.3.1 SOP Classes and Transfer Syntaxes

This Media Storage Application Profile is based on the Media Storage Service Class (see PS3.4).

Table H.3-1. STD-GEN-DVD and STD-GEN-SEC-DVD SOP Classes and Transfer Syntaxes

Information Object Definition

SOP Class UID

Transfer Syntax Name and UID

FSC Requirement

FSR Requirement

Basic Directory

1.2.840.10008.1.3.10

Explicit VR Little Endian

1.2.840.10008.1.2.1

Mandatory

Mandatory

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

Explicit VR Little Endian

1.2.840.10008.1.2.1

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG Lossless, Non-Hierarchical, First-Order Prediction (Process 14 [Selection Value 1])

1.2.840.10008.1.2.4.70

Defined in Conformance Statement

Mandatory for -JPEG profiles for all SOP Classes defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG Baseline (Process 1)

1.2.840.10008.1.2.4.50

Defined in Conformance Statement

Mandatory for -JPEG profiles for all SOP Classes defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG Extended (Process 2 & 4)

1.2.840.10008.1.2.4.51

Defined in Conformance Statement

Mandatory for -JPEG profiles for all SOP Classes defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG 2000 Image Compression (Lossless Only)

1.2.840.10008.1.2.4.90

Defined in Conformance Statement

Mandatory for -J2K profiles for all SOP Classes defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG 2000 Image Compression

1.2.840.10008.1.2.4.91

Defined in Conformance Statement

Mandatory for -J2K profiles for all SOP Classes defined in Conformance Statement


The SOP Classes and corresponding Transfer Syntaxes supported by these Media Storage Application Profiles are specified in the Table H.3-1. The supported Storage SOP Class(es) shall be listed in the Conformance Statement.

DICOM PS3.11 2026a - Media Storage Application Profiles