DICOM PS3.11 2026b - Media Storage Application Profiles

N General Purpose BD With MPEG-4 AVC/H.264 Level 4.2 Compression Interchange Profiles (Normative)

N.1 Profile Identification

This Annex defines a Media Storage Application Profile Class potentially inclusive of all defined Media Storage SOP Classes. This class is intended to be used for the interchange of Composite SOP Instances via BD media for general-purpose applications. Objects from multiple modalities may be included on the same media. Multi-frame Images and video may be compressed with MPEG-4 AVC/H.264 High Profile / Level 4.2 or MPEG-4 AVC/H.264 Stereo High Profile / Level 4.2.

A detailed list of the Media Storage SOP Classes that may be supported is defined in PS3.4.

Table N.1-1. STD-GEN-BD-MPEG4-LV42 and STD-GEN-SEC-BD-MPEG4-LV42 Profiles

Media Storage Application Profile

Identifier

Description

General Purpose BD Interchange with MPEG-4 AVC/H.264 HiP@Level4.2 for 2D video

STD-GEN-BD-MPEG4-HPLV42-2D

Handles interchange of Multi-frame Images and video using MPEG-4 AVC/H.264 HiP@Level4.2 compression for 2D video.

General Purpose BD Interchange with MPEG-4 AVC/H.264 HiP@Level4.2 for 3D video

STD-GEN-BD-MPEG4-HPLV42-3D

Handles interchange of Multi-frame Images and video using MPEG-4 AVC/H.264 HiP@Level4.2 compression for 3D video.

General Purpose BD Interchange with MPEG-4 AVC/H.264 Stereo HiP@Level4.2

STD-GEN-BD-MPEG4-SHPLV42

Handles interchange of Multi-frame Images and video using MPEG-4 AVC/H.264 Stereo High Profile / Level4.2 compression.

General Purpose Secure BD Interchange with MPEG-4 AVC/H.264 HiP@Level4.2 for 2D video

STD-GEN-SEC-BD-MPEG4-HPLV42-2D

Handles interchange of Multi-frame Images and video using MPEG-4 AVC/H.264 HiP@Level4.2 compression for 2D video. Offers confidentiality, integrity and, depending on the File-set Creator's choice, data origin authentication.

General Purpose Secure BD Interchange with MPEG-4 AVC/H.264 HiP@Level4.2 for 3D video

STD-GEN-SEC-BD-MPEG4-HPLV42-3D

Handles interchange of Multi-frame Images and video using MPEG-4 AVC/H.264 HiP@Level4.2 compression for 3D video. Offers confidentiality, integrity and, depending on the File-set Creator's choice, data origin authentication.

General Purpose Secure BD Interchange with MPEG-4 AVC/H.264 Stereo HiP @ Level4.2

STD-GEN-SEC-BD-MPEG4-SHPLV42

Handles interchange of Multi-frame Images and video using MPEG-4 AVC/H.264 Stereo High Profile / Level4.2 compression. Offers confidentiality, integrity and, depending on the File-set Creator's choice, data origin authentication.


Equipment claiming conformance to this Media Storage Application Profile shall list the subset of Media Storage SOP Classes that it supports in its Conformance Statement.

Note

Since it is not required to support all Media Storage Classes the user should carefully consider the subset of supported Media Storage SOP Classes in the Conformance Statements of such equipment to establish effective object interchange.

DICOM PS3.11 2026b - Media Storage Application Profiles