B 1024 X-Ray Angiographic Media Storage Application Profile (Normative)

B.1 Class and Profile Identification

This Annex defines a class of Media Storage Application Profiles for 1024 X-Ray Angiographic clinical applications. The identifier for this class shall be STD-XA1K. It is the intent of these profiles to be backward compatible with the Basic Cardiac X-Ray Angiographic Media Storage Application Profile (STD-XABC-CD) in Annex A.

The specific Media Storage Application Profiles in this class are shown in the Table B.1-1.

Table B.1-1. 1024 X-Ray Angiographic Profiles

Media Storage Application Profile

Identifier

Description

1024 X-Ray Angiographic Studies on CD-R Media

STD-XA1K-CD

It handles single frame or multi-frame X-Ray digital images up to 1024x1024x12 bits; biplane acquisitions are encoded as two single plane information objects. Secondary Capture images are supported.

1024 X-Ray Angiographic Studies on DVD Media

STD-XA1K-DVD

It handles single frame or multi-frame X-Ray digital images up to 1024x1024x12 bits; biplane acquisitions are encoded as two single plane information objects. Secondary Capture images are supported.