DICOM PS3.7 2022b - Message Exchange |
---|
The C-FIND-RSP Message contains fields as defined in Table 9.3-4 and in Annex C. Each field shall conform to DICOM encoding and Value Representation as defined in PS3.5. Fields are required as specified in the C-FIND service definition unless otherwise noted in Table 9.3-4. Fields not specified in the C-FIND service definition but present in Table 9.3-4 are required by the DIMSE-C protocol.
Table 9.3-4. C-FIND-RSP Message Fields
The even number of bytes from the end of the value field to the beginning of the next group. |
||||
This field distinguishes the DIMSE-C operation conveyed by this Message. The value of this field shall be set to 8020H for the C-FIND-RSP Message. |
||||
Shall be set to the value of the Message ID (0000,0110) field used in associated C-FIND-RQ Message. |
||||
This field indicates if a Data Set is present in the Message. This field shall be set to the value of 0101H (Null) if no Data Set is present; any other value indicates a Data Set is included in the Message. |
||||
The value of this field depends upon the status type. Annex C defines the encoding of the status types defined in the service definition. |
||||
A Data Set that encodes the Identifier that was matched. See Section 9.1.2.1.5. |
DICOM PS3.7 2022b - Message Exchange |
---|