DICOM PS3.7 2022b - Message Exchange |
---|
The N-CREATE-RSP Message contains fields as defined in Table 10.3-10 and Annex C. Each field shall conform to DICOM encoding and Value Representation as defined in PS3.5. Fields are required as specified in the N-CREATE service definition unless otherwise noted in Table10.3-10. Fields not specified in the N-CREATE service definition but present in Table 10.3-10 are required by the DIMSE-N protocol.
Table 10.3-10. N-CREATE-RSP Message Fields
The even number of bytes from the end of the value field to the beginning of the next group. |
||||
This field distinguishes the DIMSE-N operation conveyed by this Message. The value of this field shall be set to 8140H for the N-CREATE-RSP Message. |
||||
Shall be set to the value of the Message ID (0000,0110) field used in associated N-CREATE-RQ Message. |
||||
This field indicates if a Data Set is present in the Message. This field shall be set to the value of 0101H if no Data Set is present; any other value indicates a Data Set is included in the Message. |
||||
The value of this field depends upon the status type. Annex C defines the encoding of the status types defined in the service definition. |
||||
This field is encoded as a Data Set. One Data Element is encoded for each Attribute and Attribute Value applicable to the operation. |
DICOM PS3.7 2022b - Message Exchange |
---|