DICOM PS3.4 2022b - Service Class Specifications |
---|
An implementation that is conformant to this SOP Class as an SCP shall meet conformance requirements for the operations that it performs.
The SCP Conformance Statement shall be formatted as defined in Annex A “DICOM Conformance Statement Template (Normative)” in PS3.2.
The SCP Conformance Statement shall provide information on the behavior of the SCP (in terms of real world activities) for each reported value of Instance Availability (0008,0056).
The SCP Conformance Statement shall describe the behavioral relationship between the Instance Availability Notification and the Performed Procedure Step SOP Classes, if the latter are supported.
DICOM PS3.4 2022b - Service Class Specifications |
---|