DICOM PS3.4 2022c - Service Class Specifications |
---|
The DICOM AEs that claim conformance to this SOP Class as an SCU and/or an SCP shall support the Event Types and Event Information as specified in Table KK.2-2. Descriptions of the Attributes of the Event Information are specified in Section B.30 “Inventory Creation IOD” in PS3.3.
Within the Inventory Terminated with Instances Event Information, either Retrieve AE Title (0008,0054) or Retrieve URL (0008,1190), or both, shall be present.
File Access URI (0008,0409) provides a non-DICOM Protocol access method (see Annex P “Stored File Access Through Non-DICOM Protocols (Normative)” in PS3.3) to a produced Inventory SOP Instance encodedin accordance with the DICOM File Format (see Section 7 “DICOM File Format” in PS3.10). Stored Instance Base URI (0008,0407) shall be present if File Access URI (0008,0409) is a relative path reference URI. Stored Instance Transfer Syntax UID (0008,040E) and Container File Type (0008,040A) shall be present if File Access URI (0008,0409) is present. Filename in Container (0008,040B) shall be present if Container File Type (0008,040A) is ZIP, TAR, or TARGZIP. File Offset in Container (0008,040C) and File Length in Container (0008,040D) shall be present if Container File Type (0008,040A) is BLOB.
DICOM PS3.4 2022c - Service Class Specifications |
---|